![Fluke PG9000 Series Скачать руководство пользователя страница 58](http://html1.mh-extra.com/html/fluke/pg9000-series/pg9000-series_operation-and-maintenance-manual_2309961058.webp)
PG9602™ OPERATION AND MAINTENANCE MANUAL
© 2011 Fluke Calibration
Page
48
AMH
ERROR
NUMBER
PG TERMINAL MESSAGE
COMMENT
7
None
Missing or improper argument.
9
None
Unknown Command.
10
None
Invalid command suffix.
13
None
Receive buffer overflow. Buffer has been flushed.
18
None
Command not available.
60
None
Watchdog reset has occurred.
100
None
Present mass setting unknown.
102
None
6.2 kg binary mass not available on AMH.
103
<AMH memory failure>
Mass memory.
104
<AMH memory failure>
Non-volatile memory is corrupt.
105
<AMH logic failure>
Fatal electronics error.
106
None
Mass load must first be lifted.
107
<AMH logic failure>
Incorrect CPLD version.
108
<AMH logic failure>
CPLD logic error.
110
<AMH unable to rotate column
motor>
Column motor switch error on column 0 (C0).
111
<AMH unable to rotate column
motor>
Column motor switch error on column 1 (C1).
112
<AMH unable to rotate column
motor>
Column motor switch error on column 2 (C2).
113
<AMH unable to rotate column
motor>
Wrong column switch detected.
114
<AMH logic failure>
JTAG buffer overflow.
115
<AMH logic failure>
JTAG max retries.
116
<AMH logic failure>
JTAG TDO mismatch.
117
<AMH logic failure>
JTAG illegal state.
118
<AMH logic failure>
JTAG ENDxR state unknown.
119
<AMH logic failure>
JTAG unknown error.
120
<AMH unable to lift mass load>
Mass lift valve not sensed.
121
<AMH unable to lower mass load>
Mass is dropping unexpectedly.
122
<AMH unable to lift mass load>
Mass never lifted.
123
<AMH unable to lower mass load>
Mass is not dropping when it should.
124
<AMH unable to lift mass load>
IR beam malfunction. AMH will not attempt lift
125
<AMH unable to rotate column
motor>
Column motor switch on column 1 is misaligned.
126
<AMH unable to rotate column
motor>
Column motor switch on column 2 is misaligned.
130
<AMH valve failure>
6.4 kg binary mass pin engage valve (VLV#14) not sensed.
131
<AMH valve failure>
3.2 kg binary mass pin engage valve (VLV#6) not sensed.
132
<AMH valve failure>
1.6 kg binary mass pin engage valve (VLV#12) not sensed.
133
<AMH valve failure>
0.8 kg binary mass pin engage valve (VLV#4) not sensed.
134
<AMH valve failure>
0.4 kg binary mass pin engage valve (VLV#10) not sensed.
135
<AMH valve failure>
0.2 kg binary mass pin engage valve (VLV#2) not sensed.
136
<AMH valve failure>
0.1 kg binary mass pin engage valve (VLV#8) not sensed.
137
<AMH valve failure>
6.4 kg binary mass pin disengage valve (VLV#15) not
sensed.
138
<AMH valve failure>
3.2 kg binary mass pin disengage valve (VLV#7) not sensed.
Содержание PG9000 Series
Страница 10: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page X Notes...
Страница 128: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 118 Notes...
Страница 164: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 154 Notes...
Страница 188: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 178 Notes...
Страница 192: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 182 Notes...