OC3port Plus
User Manual
6-60
Step 2: Specify the number of cells to expect in the trigger burst.
1. Press the
ATM
key. You return to the
ATM SELECTIONS
menu.
2. Press the
Down Arrow
key to move the
Setup XMIT Streams
option to the
selection line. Then, press
ENTER
. The submenu for
TRANSMIT STREAM 1
is
displayed.
3. Press the
Right Arrow
key to display the submenu for
TRANSMIT STREAM 2.
4.
Press the
Down Arrow
key to move the
PCR
option to the selection line, as in:
TRANSMIT STREAM: 2
GFC 0
PCR:000037.44 MBPS
SCR:000000.00 MBPS
za147f.eps
5. Press the
EDIT
key and enter a value for
PCR
that matches the value you
supplied when you set up TRANSMIT STREAM 1 on the master unit. Press
ENTER
when you finish to save the change.
Note
The default form for expressing the
PCR
value is in megabits per
second (
MBS
). You can change the form to a percent (%
BW
) or cells
per second (
CPS
). For information on how to do this, see "Changing
the Form for the PCR and SCR" in Chapter 5.
6.
Press the
Down Arrow
key to move the
MBS
option to the selection line:
TRANSMIT STREAM: 2
SCR: 000000.00 MBPS
MBS: 0010 Cells
CDV LIM: 020.00 Cls
za142f.eps
7. Press the
EDIT
key and enter a value for
MBS
that matches the value you
supplied for MBS when you set up TRANSMIT STREAM 1 on the master
unit. Press
ENTER
when you finish to save the change.
The slave unit is now configured with the number of cells to expect in the
trigger burst.