Motor Drive Analyzer
Specifications
31
Harmonics and Spectrum
Harmonics ................................................................DC to 51st
Spectrum ranges ......................................................1 kHz to 9 kHz, 9 kHz to 150 kHz (20 MHz filter on), up to 500 MHz (voltage modulation)
Shaft Voltage
Events / second ........................................................Indicative percentage based on rise and fall time (Impulse discharges) measurements
Report Data Capture
Number of screens ...................................................Typical 50 screens can be saved in Reports (depends on compression ratio)
Transfer to PC ..........................................................Using 2 GB USB stick or mini-USB to USB cable and FlukeView
®
2 for ScopeMeter
®
Probe Settings
Voltage Probe ...........................................................1:1, 10:1, 100:1, 1000:1, 20:1, 200:1
Current Clamp ..........................................................0.1 mV/A, 1 mV/A, 10 mV/A, 20mV/A, 50mV/A, 100mV/A, 200 mV/A, 400 mV/A
Shaft Voltage Probe..................................................1:1, 10:1, 100:1
VPS4xx Probe accuracy when adjusted on the test tool
DC to 20 kHz .........................................................±1 %
20 kHz to 1 MHz ....................................................±2 %
1 MHz to 25 MHz...................................................±3 % (for higher frequencies the probe roll off starts affecting the accuracy)
Safety
General .....................................................................IEC 61010-1: Pollution Degree 2
Measurement
IEC 61010-2-030
BNC Input A, B, (C, D)
From any terminal to
earth ground....................................................1000 V CAT III, 600 V CAT IV
Between any terminal .....................................300 V CAT IV
IEC 61010-2-031
Voltage Probe VPS410 10:1
From any terminal to
earth ground....................................................1000 V CAT III, 600 V CAT IV
Between any terminal .....................................1000 V CAT III, 600 V CAT IV
Voltage Probe VPS42x 100:1
From any terminal to
earth ground....................................................1000 V CAT III 600 V CAT IV
Between probe tip and
reference lead .................................................2000 V
Note: Voltage ratings are given as “working voltage”. They should be read as Vac-rms (50-60 Hz) for ac sinewave applications and as Vdc for dc
applications.