Table 3-1. Memory Card Error Codes
Error
Probable
Cause
Err 1 Card
CARD ERROR- Unable to use a card (Note 1):
Card is missing or card is not fully inserted.
Install a memory card (Fig. 3-1).
Card is unformatted.
Initialize a memory card (Fig. 3-3).
Write-protect switch in the read -only position.
Reposition switch (Fig. 3-1).
Card is 100% full of data
Erase files (Fig. 3-7/3-9) or use another
card.
Err 2 FILE
FILE ERROR- Unable to open a file (Note 1):
The selected file name already exists.
Choose another file name (Fig. 3-5/3-8)
or erase files (Fig. 3-7/3-9).
All file names are assigned (SEt00, SEt99 or
dAt00 to dAt99).
Erase files (Fig. 3-7/3-9) or use another
card.
Err 3 bAd
CARD PROBLEM (Scans Saved) (Note 1):
Card exchanged during scanning is
unformatted.
Use a formatted card (Note 2).
Card exchanged during scanning is full of
data.
Use a different card (Note 2).
Replacement card has a duplicate file name.
(Note 4).
Use a different card (Note 2).
Err 4 bAd
CARD PROBLEM (Scans Lost) (Note 1):
Same as Err 3 bAd (Note 3).
Err 3 FULL
CARD IS FULL (Scans Saved) (Note 1):
Card is 100% full of data.
Use a different card (Note 2).
Err 4 FULL
CARD IS FULL (Scans Lost) (Note 1):
Sane as Err 3 FULL (Note 3)
Note 1: Err 1 and Err 2 are non-scanning errors that occur only before scanning starts. Err 3 and Err 4
are scanning errors that occur only after scanning starts.
Note 2: Err 3 indicates scans are being saved in an internal memory (75 scans) while the memory card
error is being resolved. Err 4 indicates scans are bin lost because the internal memory overflowed (75
scans) before the error was corrected. When a suitable exchange card is inserted, the internal memory
updates the card with the stored scans.
Note 3: When memory cards are exchanged during scanning and the replacement card has a problem,
Err 3/Err 4 is appended with the word bAd. When the memory card used for scanning becomes full of
data, Err 3/Err 4 is appended with the word FULL.
Note 4: When memory cards are exchanged during scanning, the replacement card must have the same
file name available as was used for the original scan. If this file name already exists on the replacement
card, and Err 3/Err 4 will occur.
Содержание 2635A
Страница 8: ...2635A Users Manual vi...
Страница 26: ...2635A Users Manual xxiv...
Страница 29: ...Preparation for Use Introduction 1 1 3...
Страница 50: ...2635A Users Manual 1 24...
Страница 53: ...Front Panel Operations Summary of Front Panel Operations 2 2 3...
Страница 88: ...2635A Users Manual 2 38...
Страница 180: ...2635A Users Manual 5 10...
Страница 212: ......
Страница 234: ...2635A Users Manual B 4...
Страница 250: ...2635A Users Manual D 8...
Страница 251: ...E 1 Appendix E 8 Bit Binary Coded Decimal Table...
Страница 264: ...Hydra Memory Card Record DATA FILES SET UP FILES DATxx Application Note SETxx Application Note...
Страница 268: ...2635A Users Manual 4...