A-1
Appendix A
Specifications
Introduction
The instrument specifications presented here are applicable within the conditions listed
in the Environmental chapter.
The specifications state total instrument accuracy following calibration, including:
•
A/D errors
•
Linearization conformity
•
Initial calibration errors
•
Isothermality errors
•
Relay thermal emf’s
•
Reference junction conformity
•
Temperature coefficients
•
Humidity errors
Sensor inaccuracies are not included in the accuracy figures.
Accuracies at Ambient Temperatures Other than Specified
To determine typical accuracies at temperatures intermediate to those listed in the
specification tables, linearly interpolate between the applicable 0ºC to 60ºC and 18ºC to
28ºC accuracy specifications.
Response Times
Refer to Typical Scanning Rate and Maximum Autoranging Time later in this Appendix.
DC Voltage Measurements
Resolution
See Table A-1