2635A
Users Manual
4-52
Table 4-5. Command and Query Reference (cont)
SCALE_MB?
Mx+B Scaling Values Query
Return the M and B scaling values for the indicated channel.
SCALE_MB?
<channel>
<channel> = 0, ... 20
This command returns three values. The first two are the M and B values for the
channel indicated, even when M=1 and B=0 or the channel function is defined as
OFF. These first two values are returned in M and B order and in scientific notation
format with five digits of resolution. The third value returned indicates the Mx+B
display range. The Mx+B scaling values are automatically reset to 1 (M) and 0 (B)
when the function for that channel is changed.
The range code for the display <range> is shown below:
Range
Display
Max B
Code
Offset
Value
1
0.0000 m
9.9999E-3
2
00.000 m
99.999E-3
3
000.00 m
999.99E-3
4
0000.0 m
9999.9E-3
5
0.0000 x1
9.9999
6
00.000 x1
99.99
7
000.00 x1
999.99
8
0000.0 x1
9999.9
9
0.0000k
9.9999E3
10
00.000k
99.999E3
11
000.00 k
999.99E3
12
0000.0 k
9999.9E3
13
0.0000 M
9.9999E6
14
00.000 M
99.999E6
15
000.00 M
999.99E6
16
0000.0 M
9999.9E6
Example: SCALE_MB? 0 r1.0000E+0,-1.0000E+3,9 [For channel 0, M=1,
B=-1000, and the display range is 0.0000 k.]
SCAN
Enable/Disable Scanning
This command performs the same function as
Q
on the front panel.
SCAN 1
Enable scanning.
SCAN 0
Disable scanning
If SCAN 0 is set during the measurement interval of the scan, the measurement
portion is completed. If SCAN 0 is set during the countdown interval of the scan, the
scan is immediately terminated.
If there are no configured channels (all are defined as OFF) or values other than 0 or
1 are given, an Execution Error is generated.
The MON and SCAN commands work in conjunction with the front panel controls.
The Monitor and Scan functions can be enabled or disabled from either the front
panel or the computer interface. The most recently specified monitor channel (from
front panel or computer interface) becomes the one channel monitored.
Front panel
Q
and
M
buttons work only when the lockout state is “local without
lockout” (see the LOCS command)..
Содержание 2635A
Страница 8: ...2635A Users Manual vi...
Страница 26: ...2635A Users Manual xxiv...
Страница 29: ...Preparation for Use Introduction 1 1 3...
Страница 50: ...2635A Users Manual 1 24...
Страница 53: ...Front Panel Operations Summary of Front Panel Operations 2 2 3...
Страница 88: ...2635A Users Manual 2 38...
Страница 180: ...2635A Users Manual 5 10...
Страница 212: ......
Страница 234: ...2635A Users Manual B 4...
Страница 250: ...2635A Users Manual D 8...
Страница 251: ...E 1 Appendix E 8 Bit Binary Coded Decimal Table...
Страница 264: ...Hydra Memory Card Record DATA FILES SET UP FILES DATxx Application Note SETxx Application Note...
Страница 268: ...2635A Users Manual 4...