2635A
Users Manual
B-2
AC Signal Cross Talk into an AC Voltage Channel
ACV Error Ratio =
VACrms error
VACrms crosstalk
Frequency crosstalk
(
)
(
)
(
)
×
Range
Ratio (worst case)
Ratio (typical)
300.00 m
4.8
×
10
-8
v
v
Hz
×
1.4
×
10
-8
v
v
Hz
×
3.0000 V
1.1
×
10
-7
v
v
Hz
×
3.0
×
10
-8
v
v
Hz
×
30.000 V
1.2
×
10
-6
v
v
Hz
×
2.6
×
10
-7
v
v
Hz
×
150.00/300.00
1.2
×
10
-5
v
v
Hz
×
3.4
×
10
-6
v
v
Hz
×
For example, the typical effect on a channel measuring ac volts (300 mV range) by a
220V ac signal at 60 Hz on another channel would be: 220 X 60 X 1.4 X 10
-8
= 0.18 mV.
AC Signal Cross Talk into an Ohms Channel
AC Frequency = 50, 60 Hz, ±0.1%
OHMS Error Ratio
1
=
Ohms error
VACrms crosstalk
(
)
(
)
Range
Ratio (worst case)
Ratio (typical)
300.00
e
3.3
×
10
-5
Ohms
Vrms
NoEffect
3.0000 k
e
2.4
×
10
-6
kOhms
Vrms
6.7
×
10
-7
kOhms
Vrms
30.000k
e
3.1
×
10
-4
kOhms
Vrms
8.4
×
10
-5
kOhms
Vrms
300.00 k
e
5.6
×
10
-3
kOhms
Vrms
3.7
×
10
-3
kOhms
Vrms
3.0000 M
e
3.8
×
10
-4
MOhms
Vrms
3.8
×
10
-5
MOhms
Vrms
Содержание 2635A
Страница 8: ...2635A Users Manual vi...
Страница 26: ...2635A Users Manual xxiv...
Страница 29: ...Preparation for Use Introduction 1 1 3...
Страница 50: ...2635A Users Manual 1 24...
Страница 53: ...Front Panel Operations Summary of Front Panel Operations 2 2 3...
Страница 88: ...2635A Users Manual 2 38...
Страница 180: ...2635A Users Manual 5 10...
Страница 212: ......
Страница 234: ...2635A Users Manual B 4...
Страница 250: ...2635A Users Manual D 8...
Страница 251: ...E 1 Appendix E 8 Bit Binary Coded Decimal Table...
Страница 264: ...Hydra Memory Card Record DATA FILES SET UP FILES DATxx Application Note SETxx Application Note...
Страница 268: ...2635A Users Manual 4...