Maintenance
Performance Tests
7
7-19
3.
Remove the Input Module from the rear of Hydra Series II. Open the Input Module
and jumper the H (high) terminal of channels 1, 2, and 3 together. Connect a test
lead to the H of channel 1. Also jumper the L (low) terminals of channel 1, 2, and 3
together. Connect a second test lead to the L of channel 1. Reinstall the Input
Module into Hydra Series II. Refer to Figure 7-4.
4.
Switch power ON.
5.
Using a digital multimeter (DMM), verify that alarm outputs 0 through 3 are in the
OFF or HIGH state. Perform this test by connecting the low or common of the
multimeter to the ground test lead and the high of the multimeter to the alarm output.
Verify a voltage greater than +3.8V dc.
6.
Connect a cable from the Output VA HI and LO connectors of the 5700A to the V
Ω
and COM connectors on the front panel of Hydra Series II. Then jumper Hydra
Series II’s V
Ω
terminal to the H (high) test lead of the Input Module and the COM
terminal to the L (low) test lead.
7.
On Hydra Series II, select the VDC function, 3V range, and assign a HI alarm limit
of +1.0000 for channels 0 through 3. Set up all other channels (4-20) to the OFF
function. Select a scan interval of 5 seconds.
8.
Set the 5700A to 0.9900 volts.
9.
Press Hydra Series II’s
Q
button. Hydra Series II should scan channels 0 through 3
every 5 seconds.
10.
Using a digital multimeter, again verify that alarm outputs 0 through 3 are in the
OFF or HIGH state.
11.
Set the 5700A to 1.1000 volts. Verify that the alarm outputs 0 through 3 are
in the ON or LOW state (measure less than +0.8V dc).
Содержание 2635A
Страница 8: ...2635A Users Manual vi...
Страница 26: ...2635A Users Manual xxiv...
Страница 29: ...Preparation for Use Introduction 1 1 3...
Страница 50: ...2635A Users Manual 1 24...
Страница 53: ...Front Panel Operations Summary of Front Panel Operations 2 2 3...
Страница 88: ...2635A Users Manual 2 38...
Страница 180: ...2635A Users Manual 5 10...
Страница 212: ......
Страница 234: ...2635A Users Manual B 4...
Страница 250: ...2635A Users Manual D 8...
Страница 251: ...E 1 Appendix E 8 Bit Binary Coded Decimal Table...
Страница 264: ...Hydra Memory Card Record DATA FILES SET UP FILES DATxx Application Note SETxx Application Note...
Страница 268: ...2635A Users Manual 4...