2635A
Users Manual
4-8
1.
Configure the Data Bucket communication parameters, as described in Figure 4-2,
for bAUd = 9600, PAR = no, CtS = OFF, and Echo = On.
2.
Turn on the PC, start the communications software, and open the COM port
configuration screen.
3.
Select the following communications parameters
Connector COM1 [Typical]
Baud Rate 9600
Data Bits 8
Stop Bits 1
Parity None
Flow ControlNone[May be called the RTS/CTS line]
4.
In Terminal, request the Data Bucket to send its identification number by entering:
*IDN? <Enter>. If *IDN? did not appear on the screen as the characters were
entered, be sure the instrument RS-232 port is configured for Echo = On (Figure 4-
2). If the wrong characters appear, there is an incompatibility in the COM port
configurations (baud rate, parity, etc.). If everything seems normal, but characters
still don’t appear, check the RS-232 connection cable (see Appendix D). When the
RS-232 link is operating correctly, the instrument returns an identification string and
execution prompt similar to the following:
FLUKE,2635A,0,Mn.n An.n Dn.n Ln.n
=>
Mn.n identifies the main software version.
An.n identifies the analog-to-digital converter software version.
Dn.n identifies the display software version.
Ln.n identifies the programmable gate-array configuration version.
5.
Other commands can be entered from the PC to gain familiarity with the instrument
command set. All commands are summarized in Table 4-4 and explained in Table 4-
5.
For example: to reset the instrument, configure channel 0 to measure volts dc using
the 300V DC scale (scale 4), send scan results to the RS-232 port, and scan once,
enter the following:
*RST [Resets the instrument (which does not affect the communication parameters)]
=>
FUNC 0, VDC, 4 [Set channel 0 to volts dc and scale 4 (300V DC)]
=>
PRINT_TYPE 0,0 [Sets the data destination as the RS-232 port, and all data]
=>
PRINT 1 [Enables data logging to the RS-232 port]*TRG [Triggers a single scan]
=>
15:17:04 07/21/94
0: 000.00 VDC
Содержание 2635A
Страница 8: ...2635A Users Manual vi...
Страница 26: ...2635A Users Manual xxiv...
Страница 29: ...Preparation for Use Introduction 1 1 3...
Страница 50: ...2635A Users Manual 1 24...
Страница 53: ...Front Panel Operations Summary of Front Panel Operations 2 2 3...
Страница 88: ...2635A Users Manual 2 38...
Страница 180: ...2635A Users Manual 5 10...
Страница 212: ......
Страница 234: ...2635A Users Manual B 4...
Страница 250: ...2635A Users Manual D 8...
Страница 251: ...E 1 Appendix E 8 Bit Binary Coded Decimal Table...
Страница 264: ...Hydra Memory Card Record DATA FILES SET UP FILES DATxx Application Note SETxx Application Note...
Страница 268: ...2635A Users Manual 4...