7 Connection
FLUXUS G831ST-HT
7.5 Temperature probe
67
UMFLUXUS_G831ST-HTV1-0EN, 2021-06-01
Connection of the temperature probe to the junction box
• Remove the blind plug for the connection of the temperature probe.
• Open the cable gland of the temperature probe. The compression part remains in the cap nut.
• Push the cable of the temperature probe through the cap nut and the compression part.
• Prepare the cable.
• Shorten the external shield and brush it back over the compression part.
• Screw the sealing ring side of the reducer into the junction box.
• Screw the basic part into the reducer.
• Insert the cable into the junction box.
• Fix the cable gland by screwing the cap nut onto the basic part.
• Connect the temperature probe to the terminals of the junction box.
Tab. 7.13: Terminal assignment (junction box)
terminal
extension cable (KL2)
temperature probe (KL1)
1
red
red
2
gray
red/blue
3
blue
white/blue
4
white
white