7 Connection
FLUXUS G831ST-HT
7.5 Temperature probe
65
UMFLUXUS_G831ST-HTV1-0EN, 2021-06-01
7.5.2
Connection with extension cable
• Remove the blind plug for the connection of the temperature probe.
• Open the cable gland of the extension cable. The compression part remains in the cap nut.
• Push the extension cable through the cap nut, the compression part, the basic part and the reducer.
• Prepare the extension cable.
• Shorten the external shield and brush it back over the compression part.
• Insert the extension cable into the housing.
• Screw the sealing ring side of the reducer into the transmitter housing.
• Screw the basic part into the reducer.
• Fix the cable gland by screwing the cap nut onto the basic part.
• Connect the connectors of the extension cable and the temperature probe.
• Connect the extension cable to the terminals of the transmitter.
7.5.3
Connection with junction box
Connection of the extension cable to the transmitter
• Remove the blind plug for the connection of the temperature probe.
• Open the cable gland of the extension cable. The compression part remains in the cap nut.
• Push the extension cable through the cap nut, the compression part, the basic part and the reducer.
• Prepare the extension cable.
• Shorten the external shield and brush it back over the compression part.
• Insert the extension cable into the housing.
• Screw the sealing ring side of the reducer into the transmitter housing.
• Screw the basic part into the reducer.
• Fix the cable gland by screwing the cap nut onto the basic part.
• Connect the extension cable to the terminals of the transmitter.
Tab. 7.11: Terminal assignment (transmitter)
terminal
temperature probe
5
red
4
red/blue
3
white/blue
6
white
Tab. 7.12: Terminal assignment (transmitter)
terminal
extension cable
5
red
4
gray
3
blue
6
white