The total volume or total mass of the fluid at the measuring point can be determined.
There are 2 totalizers, one for the positive and the other for the negative flow direction. The unit of measurement used for
totalizing corresponds to the volume or mass unit selected for the physical quantity.
The totalizer values can be displayed in the status line during the measurement.
• Press CLR. The menu item
Execute command
is displayed.
• Select the list item
Totalizers
.
• Press ENTER.
The following scroll list appears:
By pressing
or
, the totalizers for the positive and the negative flow direction can be displayed during the
measurement.
17.1.1
Number of decimal places
The values of the totalizers can be displayed with up to 11 places, e.g.,
74890046.03
. The number of decimal places
(max. 4) can be defined.
• Select the menu item
Totalizers
in the program branch
Special functions
.
• Press ENTER.
• Select
Automatic
if the number of decimal places has to be adjusted automatically.
• Press ENTER.
Low totalizer values will initially be displayed with 3 decimal places. If the values of the totalizers are higher, the number of
decimal places will be reduced.
• Select the number of decimal places.
• Press ENTER.
Measurement\Execute command\Totalizers
display
description
Start totalizers
start and stop totalizer
Reset totalizers
reset the totalizer to zero
Freeze display
display the measured value of the totalizer for several seconds
Reset error
reset the totalizer error
Stop/clear totalizers
stop and reset the totalizer to zero
Notice!
A keystroke will only influence the totalizers of the measuring channel whose measured values are currently
displayed.
Special functions\Totalizers
max. value
display
< 10
6
±0.000
...
±999999.999
< 10
7
±1000000.00
...
±9999999.99
< 10
8
±10000000.0
...
±99999999.9
< 10
10
±1000000000
...
±9999999999