15 Data logger
15.1 Configuration of the data logger
FLUXUS G831ST-HT
2021-06-01, UMFLUXUS_G831ST-HTV1-0EN
118
Storage mode
• Select the menu item
Configuration
.
• Press ENTER until the menu item
Storage mode
is displayed.
• Press ENTER.
• Select
Sample
to store the current measured value.
• Select
Average
if the average of all undamped measured values of a storage interval is to be stored.
Further parameters for storing
It can be defined whether the following parameters are to be stored together with the measured values.
• Select
Yes
to store the value. Select
No
in order not to store the value.
Special functions\Data logger\Configuration\Storage mode
Notice!
The storage mode does not affect the outputs.
Notice!
Storage mode
=
Average
The average of the physical quantity and of other quantities assigned to the measuring channel, e.g., the measured
temperature, will be calculated.
If a storage rate < 5 s is selected,
Sample
will be used.
If no average could be calculated over the complete storage interval, the value will be marked as invalid.
Tab. 15.1: Parameters for storing
display
description of the parameters
Store totalizers
values of the totalizers
Store diagnost. values
diagnostic values
Store transd. temp.
transducer temperature