9 Measurement
9.2 Measurement settings
FLUXUS G831ST-HT
2021-06-01, UMFLUXUS_G831ST-HTV1-0EN
80
9.2
Measurement settings
• Select the program branch
Options
.
• Press ENTER.
• Select the menu item
Measurement
.
• Press ENTER.
9.2.1
Selection of the physical quantity
• Select the menu item
Physical quantity
.
• Press ENTER.
9.2.2
Selection of the unit of measurement
For the selected physical quantity (except sound speed), a scroll list with the available units of measurement is displayed.
The unit of measurement which was selected previously is displayed first.
• Select the unit of measurement of the physical quantity.
• Press ENTER.
9.2.3
Input of the damping factor
Each displayed measured value is a floating average of the last x seconds, with x being the damping factor. If 0 s is
entered as damping factor, no average is calculated.
The value of 10 s is appropriate for normal flow conditions. If the values fluctuate strongly, caused by a higher dynamic
flow, a higher damping factor can be very useful.
• Enter the damping factor.
• Press ENTER.
Options
Measurement
Units of measurement
Physical quantity
Flow velocity
Unit of measurement
Volumetric flow rate
Damping
Mass flow rate
Error value delay
see annex A
Measurement\Measurement
Options\Measurement\Physical quantity
Options\Physical quantity\Volumetric flow rate
Notice!
If the physical quantity or the unit of measurement is changed, the settings of the outputs will have to be checked.
Options\Measurement\Damping