15 Data logger
15.1 Configuration of the data logger
FLUXUS G831ST-HT
2021-06-01, UMFLUXUS_G831ST-HTV1-0EN
116
Event-based starting time
If it is necessary to start the storing of measured values at
a particular event, select
Event-based
as starting time.
The event is signalized via an input or event trigger. All
configured inputs and event triggers are displayed in the
scroll list.
• Select the input or the event trigger through which the
event is to be signalized.
• Press ENTER.
The storage rate is the frequency to transmit or store
measured values.
• Select in the scroll list a storage rate for storing the measured values in case the event does not occur.
• Press ENTER.
• Select
Off
if no measured values are to be stored, as long as the event does not occur.
• Press ENTER.
• Select in the scroll list a storage rate for storing the measured values in case the event occurs.
• Press ENTER.
Example
actual time: 09:06 am
setting:
Full 10 minutes
The storing starts at 09:10 am.
Notice!
Make sure that the time of all transmitters is synchronized.
Special functions
Data logger
Configuration
…
Start storing
→ Event-based
Storage rate w/o trig.
Storage rate (trig.)
Buffer time →П
Buffer time П→
see annex A
Special functions\Data logger\
Configuration\Start storing\Event-based
Special functions\Data logger\
Configuration\Start storing\Event-based\
Storage rate w/o trig.
Special functions\Data logger\Configuration\Start storing\Event-based\Storage rate
(trig.)