7 Connection
7.5 Temperature probe
FLUXUS G831ST-HT
2021-06-01, UMFLUXUS_G831ST-HTV1-0EN
64
7.5.1
Direct connection of the temperature probe
• Remove the blind plug for the connection of the temperature probe.
• Open the cable gland of the temperature probe. The compression part remains in the cap nut.
• Push the cable of the temperature probe through the cap nut, the compression part, the basic part and the reducer.
• Prepare the cable.
• Shorten the external shield and brush it back over the compression part.
• Insert the cable into the housing.
• Screw the sealing ring side of the reducer into the transmitter housing.
• Screw the basic part into the reducer.
• Fix the cable gland by screwing the cap nut onto the basic part.
• Connect the temperature probe to the terminals of the transmitter.
Tab. 7.10: Connection systems
direct connection
connection with extension cable
connection with junction box
Fig. 7.14: Preparation of the temperature probe
1 – cap nut
2 – compression part
3 – basic part
4 – reducer
5 – external shield, brushed back
6 – sealing ring side