69
7 Installation of FLUXUS G709
R
LOAD
is the sum of all ohmic resistances in the circuit (resistance of the conductors, inner resis-
tance of the external power supply, resistance of the amperemeter/voltmeter, etc.)
semi-active
current loop,
operated as
passive cur-
rent loop
Px+
(red)
(black)
Px-
U
H
= 0...24 V
U
H
> 0.021A
.
R
LOAD
[
]
passive
current loop
Px+
(red)
(black)
Px-
U
H
= 5...25 V
U
H
>0.021A
.
R
LOAD
[
]+4V
If U
H
= 12 V, R
LOAD
must
be in the range 0...380
.
voltage
output
Px+
(red)
(black)
Px-
R
LOAD
> 2 M
If R
LOAD
is lower, the
accuracy is lower than
specified.
binary output
(open
collector)
Px+
(red)
(black)
Px-
U
H
= 5...24 V
R
C
[k
] = U
H
/I
c
[mA]
I
c
= 1...4 mA
binary output
(Reed relay)
Px+
Pxa
Pxb
Px-
U
MAX
= 24 V
I
MAX
= 150 mA
Table 7.9: Circuits of the outputs
output
FLUXUS
terminal
circuits
+
-
mA
+
-
+ -
U
H
mA
+
-
+ -
U
H
+
Ri=500
R
i
= 500
V
+
-
+
-
U
H
V
+
-
R
C
a
b
Содержание FLUXUS G704
Страница 4: ......
Страница 174: ...174 UMG70XV3 4EN 12 01 2009 FLUXUS G709 dimensions in mm 198 1 208 25 129 222 2 15 122 5 110 M2 5 43 7 213 129 170 5...
Страница 235: ...UMG70XV3 4EN 12 01 2009 235 D Certificates D Certificates...
Страница 236: ...236 UMG70XV3 4EN 12 01 2009 D Certificates...
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ...UMG70XV3 4EN 12 01 2009 D Certificates...
Страница 241: ......
Страница 242: ......
Страница 243: ......
Страница 244: ......
Страница 245: ......
Страница 246: ......
Страница 247: ......
Страница 248: ......
Страница 249: ......
Страница 250: ......
Страница 251: ......
Страница 252: ......
Страница 253: ......
Страница 254: ......
Страница 255: ......
Страница 256: ......
Страница 257: ......