18 Outputs
18.4 Activation of a Pulse Output
A pulse output is an integrating output which emits a pulse when the volume or the mass
of the medium which has passed the measuring point reaches a given value (
PULSE
VALUE
). The integrated quantity is the selected physical quantity. Integration is restarted
when a pulse is emitted.
Enter in
FULL-SCALE VAL.
the highest measured value
expected.
FULL-SCALE VAL.
is the measured value corresponding
to the upper limit of the output range as defined in section
18.1.1.
example:
The output range 4…20 mA was selected for a current loop, the zero-scale
value was set to 0 m
3
/h and the full-scale value to 300 m
3
/h.
If the volume flow is 300 m
3
/h, a 20 mA signal is transmitted to the output.
If the volume flow is 0 m
3
/h, a 4 mA signal is transmitted to the output.
Note!
The menu item
PULSE OUTPUT
will be displayed in the program
branch
OUTPUT OPTIONS
only if a pulse output has been installed.
Select in the program branch
OUTPUT OPTIONS
the chan-
nel for which a pulse output has to be activated. Press EN-
TER.
This display will not be indicated, if the flowmeter has only
one measuring channel.
Select
YES
to activate the output. Press ENTER.
This error message will be displayed if the flow velocity is
selected as physical quantity.
The use of the pulse output is not possible in this case as
integration of the flow velocity does not result in a reason-
able value.
Enter the
PULSE VALUE
. The unit of measurement will be
displayed according to the current physical quantity.
When the totalized physical quantity reaches the pulse val-
ue, a pulse will be emitted.
Enter the
PULSE WIDTH
.
The range of possible pulse widths depends on the specifi-
cation of the instrument (e.g. counter, PLC) to be connect-
ed with the output.
Full-Scale Val.
300.00
m3/h
Output Options
for Channel
A:
Pulse Output
B1: no
>YES<
Pulse Output
NO COUNTING
!
Pulse Value
0.01
m3
Pulse Width
100
ms
Содержание FLUXUS G704
Страница 4: ......
Страница 174: ...174 UMG70XV3 4EN 12 01 2009 FLUXUS G709 dimensions in mm 198 1 208 25 129 222 2 15 122 5 110 M2 5 43 7 213 129 170 5...
Страница 235: ...UMG70XV3 4EN 12 01 2009 235 D Certificates D Certificates...
Страница 236: ...236 UMG70XV3 4EN 12 01 2009 D Certificates...
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ...UMG70XV3 4EN 12 01 2009 D Certificates...
Страница 241: ......
Страница 242: ......
Страница 243: ......
Страница 244: ......
Страница 245: ......
Страница 246: ......
Страница 247: ......
Страница 248: ......
Страница 249: ......
Страница 250: ......
Страница 251: ......
Страница 252: ......
Страница 253: ......
Страница 254: ......
Страница 255: ......
Страница 256: ......
Страница 257: ......