175
A Technical Data
Transducers
Shear Wave Transducers (for ATEX zone 1)
technical type
GDK1N31
GDM1N31
GDG1N31
order code
GSK-NA1TS
GSM-NA1TS
GSG-NA1TS
transducer frequency MHz 0.5
1
0.2
medium pressure
1
min. extended
bar
20
20
20
min.
bar
metal pipe: 30
plastic pipe: 1
metal pipe: 30
plastic pipe: 1
metal pipe: 30
plastic pipe: 1
outer pipe diameter
2
min. extended
mm
70
30
250
min. recommended
mm
80
40
380
max. recommended
mm
500
80
810
max. extended
mm
720
120
1100
pipe wall thickness
min.
mm
5
2.5
14
max.
mm
-
-
-
material
housing PEEK
with
stainless steel
cap304
(1.4301)
stainless steel
316 Ti (1.4571)
PEEK with
stainless steel
cap304
(1.4301)
contact surface
PEEK
PEEK
PEEK
degree of protection
according to EN 60529
IP 65
IP 65
IP 65
transducer cable
type 2549
2549
2549
length m
5
4
5
dimensions
length l
mm
126.5
60
129.5
width b
mm
50
30
50
height h
mm
53.5
33.5
64
dimensional drawing
operating temperature
min.
°C
-40
-40
-40
max.
°C
+130
+130
+130
continued on next page
h
b
l
h
b
l
h
b
l
Содержание FLUXUS G704
Страница 4: ......
Страница 174: ...174 UMG70XV3 4EN 12 01 2009 FLUXUS G709 dimensions in mm 198 1 208 25 129 222 2 15 122 5 110 M2 5 43 7 213 129 170 5...
Страница 235: ...UMG70XV3 4EN 12 01 2009 235 D Certificates D Certificates...
Страница 236: ...236 UMG70XV3 4EN 12 01 2009 D Certificates...
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ...UMG70XV3 4EN 12 01 2009 D Certificates...
Страница 241: ......
Страница 242: ......
Страница 243: ......
Страница 244: ......
Страница 245: ......
Страница 246: ......
Страница 247: ......
Страница 248: ......
Страница 249: ......
Страница 250: ......
Страница 251: ......
Страница 252: ......
Страница 253: ......
Страница 254: ......
Страница 255: ......
Страница 256: ......
Страница 257: ......