152
18 Outputs
18
Outputs
If the flowmeter is equipped with outputs, they have to be installed and activated before
they can be used:
• assigning a measuring channel (source channel) to the output (if the flowmeter has
more than one measuring channel)
• assigning the physical quantity (source item) to be transmitted to the output by the
source channel, and the properties of the signal
• defining the behavior of the output in case no valid measured value is available
• activation of the installed output in the program branch
OUTPUT OPTIONS
18.1 Installation of an Output
The outputs will be installed in the program branch
SPECIAL FUNCTION\SYSTEM SET-
TINGS\PROC.
OUTPUTS
.
Note!
The configuration of an output will be stored at the end of the dialog.
If the dialog is quit by pressing key BRK, the changes will not be
stored.
Select
SPECIAL FUNCTION\SYSTEM SETTINGS\PROC.
OUTPUTS
. Press ENTER.
Select the output to be installed. The scroll list contains all
actually available outputs.
A tick
after a list item indicates that this output has al-
ready been installed. Press ENTER.
This display is indicated if the output has not been installed
yet. Select
YES
. Press ENTER.
If the output is already installed, select
NO
to reconfigure it
or
YES
to uninstall the output and to return to the prece-
dent menu item to select another output. Press ENTER.
Select in the scroll list the measuring channel to be as-
signed as source channel to the output. Press ENTER.
This display will not be indicated, if the flowmeter has only
one measuring channel or only one measuring channel is
active.
Select the physical quantity (source item) to be transmitted
to the output by the source channel.
If a binary output is configured, only the list items
LIMIT
and
IMPULS
will be displayed.
SYSTEM settings
Proc. outputs
Install Output
Current I1
I1 enable
no
>YES<
I1 disable
>NO<
yes
I1 Source chan.
Channel A
I1 Source item
Measured value
Содержание FLUXUS G704
Страница 4: ......
Страница 174: ...174 UMG70XV3 4EN 12 01 2009 FLUXUS G709 dimensions in mm 198 1 208 25 129 222 2 15 122 5 110 M2 5 43 7 213 129 170 5...
Страница 235: ...UMG70XV3 4EN 12 01 2009 235 D Certificates D Certificates...
Страница 236: ...236 UMG70XV3 4EN 12 01 2009 D Certificates...
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ...UMG70XV3 4EN 12 01 2009 D Certificates...
Страница 241: ......
Страница 242: ......
Страница 243: ......
Страница 244: ......
Страница 245: ......
Страница 246: ......
Страница 247: ......
Страница 248: ......
Страница 249: ......
Страница 250: ......
Страница 251: ......
Страница 252: ......
Страница 253: ......
Страница 254: ......
Страница 255: ......
Страница 256: ......
Страница 257: ......