6 Installation of FLUXUS G704
6.5
Connection of the Transducers - Connection System AS
• Remove the left blind plug for the connection of the transducers (see Fig. 6.10).
• Insert the extension cable with the AMP-Quick and SMB connectors in the housing.
• Fix the extension cable by tightening the cable gland.
• Connect the AMP-Quick and the SMB connectors with the sockets of the flowmeter
(see Fig. 6.10 and Tabelle 6.9).
Fig. 6.10: Transducers (ATEX zone 2, FM Div. 2, without explosion protection) -
direct connection
Tabelle 6.9: Terminal assignment
terminal
connection
X_AV
SMB connector (white or marketd cable)
X_AR
SMB connector (brown cable)
X1
AMP-Quick connector
AMP-Quick and
SMB connectors
cable gland
transducers:
• ATEX zone 2
• FM Div. 2
• without explosion protection
Содержание FLUXUS G704
Страница 4: ......
Страница 174: ...174 UMG70XV3 4EN 12 01 2009 FLUXUS G709 dimensions in mm 198 1 208 25 129 222 2 15 122 5 110 M2 5 43 7 213 129 170 5...
Страница 235: ...UMG70XV3 4EN 12 01 2009 235 D Certificates D Certificates...
Страница 236: ...236 UMG70XV3 4EN 12 01 2009 D Certificates...
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ...UMG70XV3 4EN 12 01 2009 D Certificates...
Страница 241: ......
Страница 242: ......
Страница 243: ......
Страница 244: ......
Страница 245: ......
Страница 246: ......
Страница 247: ......
Страница 248: ......
Страница 249: ......
Страница 250: ......
Страница 251: ......
Страница 252: ......
Страница 253: ......
Страница 254: ......
Страница 255: ......
Страница 256: ......
Страница 257: ......