Optional:
StageFlexer
and StageFlexer Jr.
devices allow strain application to cells while viewing in real
time under an upright microscope.
•
StageFlexer
®
provides equibiaxial strain to the StageFlexer
®
membrane.
•
StageFlexer Jr. provides equibiaxial or uniaxial strain to membranes cut-out of BioFlex
®
, UniFlex
®
, or Tissue
Train
®
culture plates.
F
LEX
J
R
.
™
T
ENSION
C
OMPONENTS AND
S
PECIFICATIONS
•
FlexSoft
®
Flex Jr.
™
V2.0 (Flexcell
Flex Jr.
Software) and Microsoft Windows 10
•
Flex Jr.
™
Tension FlexLink
controller
•
Flexcell
BioFlex
®
single plate baseplate
•
Gasket
•
25 mm diameter cylindrical Loading Station
™
•
Cell Seeder Loading Station
•
Silicone lubricant
•
Water trap
•
System drying filter
•
20 ft clear blue tubing (¼” (6.4 mm) O.D.) for
FLEX IN
connection
•
20 ft clear blue tubing (⅜” (9.5 mm) O.D.) for
FLEX OUT
connection
•
20 ft green tubing (⅜” (9.5 mm) O.D.) for
SYSTEM
(vacuum source) connection
•
FLEX IN moisture indicator
wye-shaped adaptor
•
Flexcell
Dynamic Culture System: Flex Jr.
™
Tension User Manual, Culture Plate & Loading Station
™
User
manual, Vacuum Source Connection User Manual, Water Trap and Drying Filter Tech Report and other
ancillary manuals for computer
•
Surge protected power outlet strip
F
LEX
J
R
.
™
T
ENSION
C
OMPUTER
S
PECIFICATIONS
•
Model:
Acer Aspire 315-51-57Z4
•
Display:
15.6" HD (1920 x 1080) resolution
•
Processor:
Intel i5 Core 7200U (2.5 GHz)
•
RAM:
6 GB DDR4 SDRAM
•
Hard Drive:
1.0 TB HDD
Operating System:
•
Microsoft Windows 10 Home Edition 64-bit
F
LEX
J
R
.
™
T
ENSION
F
LEX
L
INK
C
OMPONENT
S
PECIFICATIONS
•
Custom design vacuum controller board
o
Microprocessor sampling rate for display data transducer: 200 Hz
o
Microprocessor sampling rate for feedback/error (valve adjustment) transducer: 1 KHz
•
Single proportional valve
•
Ethernet port (RJ45)
F
LEX
J
R
.
™
T
ENSION
F
LEX
L
INK
®
C
ALIBRATION
S
PECIFICATIONS
Unconstrained Static Performance (low side transducers):
Maximum allowable waveform variance from static line: ± 0.25% elongation
Maximum % elongation error (actual vs. programmed): +/-0.6% elongation
Maximum allowable transducer inaccuracy: +/-1.0 kPa
Unconstrained Dynamic Performance (low side transducers):
Maximum % elongation error (actual vs. programmed): +/- 0.5% elongation, at the following frequencies:
<5% elongation, 0.5 and 1.0 Hz; 10 % elongation, 0.5 and 0.75Hz; >10% elongation, 0.5 Hz