17. System Message & Error Code
17.1 System Message Table
Message
Description
--oL--
Overload (Gross weight is more than Max plus 9d).
bAt.Lo
Battery level low.
HALT
Major system error detected. Power off instrument and remove power adaptor
immediately. Then check load cell connection and system power status.
L.i = xxx
•
Remain system memory is = xxx. System will start overwriting historical
transaction records (on FIFO basic) when system memory is completely full. To
preserve historical data in system memory, output all transaction records in
system memory to external backup file asap.
•
To clear this message, press
[Print/M+]
followed by
[CE/x10]
.
UndEr
Negative Weight value exceeds display range.
Reboot
Important parameters have been changed. Power off and then power on instrument
again to reboot.
------
Negative Tare value exceeds display range.
17.2 Error Code Table
Error Code
Description
Err 1
Time value error
Err 2
Date value error
Err 3
Exceed manual zero
Err 4
Offset out of range / unstable during power on
Err 5
No load cell detected
Err 6
Tare operation error
Err 7
Logic error. Hi limit set is lower than Lo limit (and Hi is not = 0)
Err 8
Logic error. Lo limit set is higher than Hi limit (and Hi is not = 0)
Err 13
Exceed maximum power on
Err 19
Capacity or division setting error (Division set is higher than 10000d)
Err 22
Manual Zero and Tare stability error
Err 23
Capacity setting error, Capacity 1 > Capacity 2
Err 24
Division setting error, d
1
> d
2
Err 25
Span gain is too low
Err 26
Not able to obtain stable status for longer than 10 seconds
Err 27
Calculated value per e of Cal 2 varies more than 1% as of Cal 1. Properly a load
cell problem.
Mass value of Cal 2 is less than 150% of Cal 1.