7. Recommended Setting Procedures for User
1.
Power off instrument and remove power adaptor.
2.
Complete connection with external platform (P2).
3.
Complete all hardware connection with peripheral.
4.
Apply power adaptor and power on.
5.
Go to F1 and check internal count value. Verify connection with weighing platform(s) is correctly done,
by applying and removing loads from weighing platform. Internal count value increases when extra
loads are applied to weighing platform. Likewise, internal count value decreases when loads are
removed from weighing platform.
6.
Go to F2 and check and make sure no display segments or indicators are missing.
7.
Go to F3 and make sure that capacity x division are correctly set.
8.
Go to F4 to check and set date format and value.
9.
Go to F5 to check and set time value.
10.
Go to F6 to do system initialization.
11.
Check and set all other operation parameters from F7, F8, F9…. F36 according to numeric order of
the internal function number.
12.
Conduct test runs to determine if the above settings best suit the application requirements. If not, set
again the corresponding parameters.
Notes
: -
a. Some of the above internal functions are platform sensitive and it is possible to have different settings
for P1 and P2 under the same internal function number.
b. Complete first all settings/checking for P1. Then go to F1 again, press
[PLU]
to change to P2. Repeat
above point #5 ~ #12 for P2.
Always contact your dealer for product information, after sales service and questions when in doubt.
8. UART & Comport Settings
This instrument is equipped with 4 built-in UARTs. Refer to below table for transmission type and assigned
comport number.
8.1 UART Transmission Type & Assigned Comport No. Table
UART #
Transmission Type
Connected to Comport #
Remark
UART 0
RS232
Comport 0
Standard
UART 1
TTL Output (not used)
RS485 and External Input Port
Not in use
RS485
Optional
UART 2
RS232
Comport 2
Standard
Bluetooth
N.A.
Optional
UART3
WIFI
N.A.
Optional
LAN
Contact dealer for support
Optional
Note
: - only one transmission type can be selected per each UART.