Setting procedures
4
: -
1.
Go to F16 and press
[Print/M+]
.
2.
Name of the current UART # appears on
LCD2
.
3.
Press
[PF/Set]
until the UART # (UART 0 ~ 3) appears on
LCD2
and press
[Print/M+]
.
4.
Press
[PF/Set]
or
[MR]
until the preferred transmission type appears on
LCD3
and press
[Print/M+]
to
set data communication types. 7 parameters are available: -
•
Off = Disabled.
•
Auto 1 = Auto print when weight is stable.
•
Auto 2 = The highest stable weight value (of a weighing process) will be printed when all loads are
removed (and gross weight returns to zero or minus).
•
Auto 3 = The last stable weight value (of a weighing process) will be printed when all loads are
removed (and gross weight returns to zero or minus).
•
Manual = Manual output by pressing
[Print/M+]
.
•
PC = Continuous or selected periodic output of real time result according to predefined format.
•
CMD = Command communication with computer and APP mode.
•
Scaner = Connect with a barcode scanner. (UART #0 and UART #2 only)
Notes: -
•
If PC or CMD or Barcode Scanner is set, refer to paragraph
8.2
for rest of the setting.
•
If Auto 1 ~ 3 or Manual is set, refer to paragraph
8.3
for rest setting.
5.
Repeat above steps 1 ~ 4 for all other UARTs.
8.2 Parameters Table for PC & Command & Scanner
Step
No.
Set Item
Parameters
PC
CMD
SCANER
1
Baud Rate
1200 ~ 921600 in 12 steps
Default = 115200
■
■
■
2
Parity
None / Odd / Even
Default = None
■
■
■
3
Data Length
7 / 8
Default = 8
■
■
■
4
Protocol Type
Note G
Protocol 0 ~ 9 & Custom
Default = Protocol 1
■
Completed
Completed
5
Transmission
Interval
Note A
0 ~ 300 (second) in 10 steps
Default = 0
Completed
8.3 Parameters Table for Auto 1 ~ 3 & Manual
5
Step
No.
Set Item
Parameters
Auto 1
Auto 2
Auto 3
Manual
1
Baud Rate
1200 ~ 921600 in 12 steps
Default = 115200
■
■
■
■
2
Parity
None / Odd / Even
Default = None
■
■
■
■
3
Data Length
7 / 8
Default = 8
■
■
■
■
4
Auto
Off / On
■
■
■
■
4
Auto 1 ~ 3 do not support PA.
5
Auto 1 ~ 3 do not support PA.