20
d.
Then press
[Zero]
to quit or
[Print/M+]
to save and continue with AD conversion speed setting.
5 AD conversion speed parameters (15 ~ 120 time per second) are available.
Press
[PF/Set]
or
[MR]
to select: -
a.
** 15
times per second.
b.
30 times per second. Recommended maximum resolution = 30,000 division
c.
60 times per second. Recommended maximum resolution = 15,000 division
d.
120 times per second. Recommended maximum resolution = 7,500 division
Note
: - After the AD conversion speed is changed, instrument will restart automatically.
F12
Auto Tare Function
** Off
On
Contin
Notes: -
•
Off
= Auto Tare Function disable.
•
On
= Only the first table weight applied will be tared off. Minimum tare load ≥ 2d.
•
Contin
= All stable weight applied will be tared off. Minimum tare load ≥ 20d.
•
If Contin is set, select also delay time (0.0 ~ 9.9 second). Delay time is the time duration from a stable weight is detected and until it
is automatically tare off.
•
Delay time value is displayed on LCD3. Enter the preferred delay time value through numeric keys and press
[Print/M+]
to save.
F13
Repetitive Tare Function
Off
** On
If F12 is set = Contin, Repetitive Tare setting “Off” will be surpassed.
F14
Operation Buzzer
Kb (keypad buzzer)
(
**On
/Off)
St (System buzzer)
(
**On
/Off)
F15
Check Buzzer & Action on
Negative Value
oFF
** in
out
Hi
Lo
oFF
= Check Buzzer disabled.
in
= Check Buzzer activates when reading is within range.
out
= Check Buzzer activates when reading is out of range.
Hi
= Check Buzzer activates when reading more than Hi limit.
Lo
= Check Buzzer activates when reading lower than Lo limit.
Action on Negative Value