Filtration Learning System
Maintenance and troubleshooting
© Festo Didactic
696684
59
12.2
Error in the piping system and tanks
The following table shows errors that can be simulated using the manual valves or that occur due to
incorrect operation.
Location
Error /
Disturbance variable /
Danger potential
What is the effect of the
error, disturbance
variable, or danger?
What causes errors?
How can I detect and
prevent errors?
Error in piping system and tanks
Waste water tank VA1
Overflow
When filling from the
upstream station, the
overflow protection is
not effective
Float switch without
function or blocked
The float switch is not
movable
Manual valve VV2
Supply to the
downstream station
No supply to
downstream station
available during
pumping
Valve VV2 is closed
Visual inspection
Manual valve VV1
Inlet for waste water
tank VA1
No supply to the storage
tank during filling
Valve VV1 is closed
Visual inspection
12.3
Error in the electrical system
The following table shows errors that can be simulated using the electrical components or that can occur
due to incorrect operation.
Location
Error /
Disturbance variable /
Danger potential
What is the effect of the
error, disturbance
variable, or danger?
What causes errors?
How can I detect and
prevent errors?
Electrical faults
19" frame
back/underside
Power supply faulty
No electrical function
Unplug the IEC power
cable or remove the
micro fuse
Observe the light at the
power supply switch or
measure the voltage
I/O terminal
No switching function
24 VDC outputs do not
switch, sensors are not
displayed by LEDs
Unplug the SysLink cable
from the I/O terminal
Visual inspection
Pump PL1
Delivery rate too low
Flow rate too low
On the motor controller,
exchange ports 1 and 2
of the pump
Visual inspection
Proportional valve VC1
Valve is not activated
Air is not blown into the
filter
2W solenoid valve –QM4
is not activated
Visual inspection
SysLink I/O terminal
Signal statuses faulty
Switching signals are not
displayed or are
displayed incorrectly
Slide switch is in
position NPN instead of
PNP
Visual inspection
with data sheet, observe
representation