4. MAINTAINING THE EQUIPMENT
Preventive maintenance
feas
ELECTRÓNICA
recommended that check the device by qualified and authorized personnel, at least every 12 months. We
recommend contacting our Customer Service for the maintenance.
feas
ELECTRÓNICA
recommended periodic inspections of the power cords, patient ECG cable, SpO
2
sensor, defibrillation
electrodes cables and pacer electrodes cables, looking for any breakage or cracks in the wires or conductors insulation.
Cleaning time is a good opportunity for a General Inspection.
feas
ELECTRÓNICA
recommends functional testing every time you start a new work shift.
Corrective Maintenance
When necessary repair the equipment, request the assistance of qualified and authorised by
feas
ELECTRÓNICA
to preserve the
warranty conditions and electrical safety parameters.
In case to request assistance from Customer Service, will be asked the device serial number, for a faster and efficient assistance.
In case of sending the device to
feas
ELECTRÓNICA
, if possible, use the original packaging. If is not possible, protect the device so
well as possible.
4.1. GENERAL INSPECTION
Periodically, inspect the defibrillator case, the power cord and paddles cables, paying special attention to sectors of the cables near the
connectors and paddles, as these are the sectors most likely to break due to rotations, tractions and bending at these points.
At any sign of deterioration require assistance to
feas
ELECTRÓNICA
’s Customer Service.
Cleaning time is a good opportunity for a General Inspection.
WARNING! There is a risk of electrical shock and death. Do not use this equipment if you doubt of the integrity of any
cable. Check the cables periodically (disconnect them before) to verify their integrity; paying special attention to cable points
close to connectors and paddles. In case you find an irregularity, request the part to our Customer’s Service.
WARNING! It is not allowed to modify the equipment.
WARNING! Not modify this equipment without authorization from the manufacturer.
WARNING! If you modify this equipment must perform appropriate tests and inspections to ensure continued safe use of
equipment.
ATTENTION! It suggests an annual contrasting against calibrated simulators.
4.2. CLEANING
WARNING! Before any cleaning, make sure the device is turned off. REMEMBER: The device has internal battery, so even
disconnected from the AC line, may be turn on. Verify the selection switch is in the off position and the display doesn’t contain
instructions and is turned off.
feas
ELECTRÓNICA
recommended clean the device and its accessories in periods of 3 months or shorter, when the dirt or the
contamination were evident.
For procedures for cleaning, disinfection and sterilization, see below.
To clean the device or its accessories, use soap neutral water to wet a piece of clean cotton cloth, wipe the device and then dry with a
cotton cloth, soft and dry.
Keep the device case free of dirt and dust.
DO NOT allow liquids to enter the device, and do not place liquids in the device while cleaning. DO NOT use abrasive cleaners or strong
solvents such as acetone or acetone-based components.
Be careful when cleaning the monitor display, is particularly sensitive to rough surfaces and susceptible to scratched.
WARNING! Risk of explosion or fire! Do not spill water or liquid on the device. Unplug the power cord and paddle
before cleaning or disinfecting the device.
ATTENTION! Do not clean the outer casing, cables or paddles with solvents, abrasives or acids.
ATTENTION! Do not clean or disinfect accessory cables, accessories and parts of the equipment or the main body
thereof with sodium hypochlorite (bleach water), solvents, acids or abrasive products. For cleaning and disinfection of
the equipment and its accessories, follow the instructions in this manual.
4.2.1. PADDLE CLEANING
NOT sterilize the external paddles with vapor or gas (ethylene oxide (ETO)), do not immerse in liquids or heat above 50ºC.
Clean the paddles, including its handle, with a solution of neutral soap and disinfect them by following the method described in section
“DISINFECTION METHODS”.
If metal parts starts to be oxidized, clean them with a soft abrasive (toothpaste can clean them well). DO NOT use strong abrasive
(steel wool).