DEFIBRILLATOR MONITOR
Mod. 3850B-BIPHASIC - USER MANUAL
Page 15 of 63
1.3. FRONT PANEL
1.3.1. MONITOR SECTION
- Connector for patient ECG cable.
- Liquid Crystal Display with backlight.
- CURSOR keys that let you use the monitor menu.
- RECORDER
key: To start or stop the paper record (just the device that has the optional recorder).
- SILENCE
key: To cancel the alarm sound for 2 minutes. Silence for two minutes is commonly used to move the patient,
reposition or change sensors and/or electrodes or any activity that involves moving the patient or temporarily disconnected. Keep the
patient under close surveillance during periods of off alarm signals. If the alarm signals are selected, the visual alarm signal will continue
to indicate the alarm condition if it occurs even though the sound is muted.
1.3.2. DEFIBRILLATOR SECTION
-
SELECTOR
SWITCH: The selector switch allows the device turn on or turn off, selecting mode MONITOR (which doesn’t
allow energy charge) and selecting the desired energy.
- CHARGE
Key: Initiates energy charge.
- DISARM
Key: disarm energy charge.
- SYNC
key: For SYNCHRONOUS mode selection.
- The shock is performed using the
keys placed in the paddles (you must press both them simultaneously) or from the front panel with
the
key.
1.3.3. RECORDER SECTION
- Cover.
- Cover open button.