DEFIBRILLATOR MONITOR
Mod. 3850B-BIPHASIC - USER MANUAL
Page 3 of 63
¾
Do not place the equipment on the patient or where it can fall over the patient. Place it next to the patient where it’s comfortable for
its use.
¾
Never attempt to introduce sharp, metallic or other objects into any aperture on the equipment.
CAUTION! The followings are the general descriptions of precautions and unsafe usage that could cause slight injuries, damage or
wrong performance of the equipment.
¾
The operation of the equipment can be affected by the presence of CAT scanner.
¾
The operation of the equipment can be affected by the presence of portable and mobile RF communications equipment.
¾
Do not use this equipment near Magnetic resonance imaging equipment (MR o MRI).
¾
The operation of the equipment below the specified amplitudes can cause inaccurate results.
¾
The Defibrillator should not be used adjacent or stacked with other equipment, if it is necessary to use it adjacent or stacked with
other equipment, the Defibrillator should be observed to verify its normal functioning in the configuration in which it is being used.
¾
To attach wires and sensors always use hypoallergenic tape.
¾
Do not store the equipment in deposits or between periods of use in places where the sun shine directly on it. Risk of damage to
the cover of the equipment, parts and accessories.
¾
Avoid installing this equipment in those places where the sun hits directly.
¾
Do not place heavy objects on the equipment.
¾
Do not drop the equipment when moving it.
¾
Use the equipment on a flat and stable surface.
¾
Important!! If using a bracket, make sure that the bracket holds at least twice the weight of the equipment. If you have any
questions, please contact
feas
ELECTRÓNICA
.
¾
Do not push the keys of the frontal panel with pushing or slicing elements. This will produce permanent damage to the keypad.
Only push the keys of the frontal panel with your fingers. Do not press the buttons with your nails.
ACCESSORIES MESSAGES
WARNINGS
•
The proper operation of the equipment and protection against the effects of the discharge of a cardiac defibrillator requires the
use of original accessories intended for this equipment. Only use original accessories provided with the equipment or those
accessories specially indicated for this equipment.
•
The use of accessories, transducers and cables other than those specified, with the exception of the transducers and cables sold
by the equipment manufacturer as replaceable parts of internal components, may cause an increase in emissions or a decrease.
•
The user shall be responsible to check the compatibility between the accessories used and this medical device.
CAUTIONS
•
Do not clean or disinfect the accessory’s cable, accessories, and parts of the equipment or its main body with sodium
hypochlorite, solvents, acids or abrasive products. For cleaning and disinfection of equipment follow the instructions given in this
manual.
•
Risk of equipment breakdown. Do not sterilize this equipment or its parts or accessories in autoclave or ethylene oxide. Do not
submerge any part of this equipment in water or other liquids or use abrasive cleaners. Do not spray or spill liquids on the
equipment or its accessories. Do not allow any liquid entering the connectors or other openings of the housing. If you
accidentally spill liquid on the equipment, turn the energy switch OFF (because the equipment has internal battery) and
disconnect it the power line (in case it is connected to the power line), clean it and dry it before reuse. If in doubt about
equipment security, send it to an authorized technical service.
BATTERY MESSAGES
WARNINGS
•
The battery must be replaced by an original battery for this equipment. Replacement with another type of battery can result in an
unacceptable risk of temperature rise, fire and/or explosion of the battery or the equipment.
CAUTIONS
•
The equipment has rechargeable battery type must remain connected to the power line during periods when it is not used.
•
If the capacity of the internal battery is below 80% the equipment will be able to function from line power or Ext +12Vdc.
•
Do not discharge the battery completely.
•
Recharge the battery immediately after it’s been used.
•
The internal battery of this equipment cannot be replaced by the user. It has to be replaced by qualified and authorized
personnel.
•
In case you replace the battery follow the local instructions to dispose Ni-Mh batteries or send them to
feas
ELECTRÓNICA
for disposal.
•
When the device is stored in warehouse, should be put to charge at least once every 60 days for at least 3hrs at 25ºC ±3ºC
temperature, to prevent battery damage.