DEFIBRILLATOR MONITOR
Mod. 3850B-BIPHASIC - USER MANUAL
Page 54 of 63
Bradycardia: 6.13s ± 0.1s.
Average HR alarm system delay
Tachycardia: 3.78s ± 0.1s.
Bradycardia: 6.09s ± 0.1s.
Maximum T-wave amplitude rejected: 1.2 times the amplitude of the QRS complex.
Precision and response of the heart rate monitor with irregular rhythms:
Signal A1: Ventricular Bigeminy
Measured and indicated frequency on display, after 20s of establishment: 80
[1/min]
± 5
[1/min]
.
Signal A2: Slow alternating Ventricular Bigeminy
Measured and indicated frequency on display, after 20s of establishment: 60
[1/min]
± 5
[1/min]
.
Signal A3: Fast alternating ventricular bigeminy
Measured and indicated frequency on display, after 20s of establishment: 120
[1/min]
± 5
[1/min]
.
Signal A4: Bidirectional Systole
Measured and indicated frequency on display, after 20s of establishment: 90
[1/min]
± 5
[1/min]
.
Maximum timing: