wavelength of red light is 660nm and 940nm for infrared light. The optical power delivered to the patient is 4mW (milliwatts).
ALARM WARNINGS
•
Because the patient is monitored but not attended continuously by an operator, it is for this reason that the alarms must be
configured and adjusted appropriately.
•
If an alarm limit is set to “---” (inhibited), the alarm condition for that limit will not be detected neither manifested by the alarm
signals (visual and auditory).
•
According to the intended use of the device, the user is considered a medical professional trained in the operation of the
device and knowledge, among others, in the monitoring and surveillance of vital signs, so any adjustment made alarm limits by
an operator (paramedics or nurse) should be carried out under the direct supervision and user commandment.
MAINTENANCE MESSAGES
WARNINGS
•
Before any cleaning, make sure the device is turned off. REMEMBER: The device has internal battery, so even disconnected
from the AC line, may be turn on. Verify the selection switch is in the OFF position and the display doesn’t contain instructions
and is turned off.
•
Risk of explosion or fire! Do not allow spill water or other liquid on the device. Unplug the power cord before cleaning or
disinfecting equipment.
•
It is not allowed to modify the equipment.
•
Not modify this equipment without authorization from the manufacturer.
•
If you modify this equipment must perform appropriate tests and inspections to ensure continued safe use of equipment.
CAUTIONS
•
Risk of equipment breakdown. Do not sterilize this equipment or its parts or accessories in autoclave or ethylene oxide. Do not
submerge any part of this equipment in water or other liquids or use abrasive cleaners. Do not spray or spill liquids on the
equipment or its accessories. Do not allow any liquid entering the connectors or other openings of the housing. If you
accidentally spill liquid on the equipment, turn the energy switch OFF (because the equipment has internal battery) and
disconnect it the power line (in case it is connected to the power line), clean it and dry it before reuse. If in doubt about
equipment security, send the same to an authorized technical service.
•
Risk of equipment breakdown. No sterilize defibrillation paddles of this device in autoclave or ethylene oxide.
•
Do not clean the external cover, the cables or the paddles with abrasive or acid products.
•
Do not clean or disinfect accessory cables, accessories and parts of the equipment or the main body thereof with sodium
hypochlorite (bleach water), solvents, acids or abrasive products. For cleaning and disinfection of the equipment and its
accessories, follow the instructions in this manual.
•
During storage in warehouses and between uses, respects the conditions of temperature, pressure and humidity as defined in
this manual and periods of recharge the internal battery specified.
•
It suggests an annual contrasting against calibrated simulators.
•
In case of fuses are damage replace with fuses of the same type and value. If the failure persists, please contact our Customer
Service.
•
This equipment has line fuses in both the neutral pole and Phase Line.
•
Neither this equipment nor any of its parts are sterile or sterilizable.
•
The user shall be responsible to check the compatibility between the accessories used and this medical device.
•
When the device is stored in warehouse, should be put to charge at least once every 60 days for at least 3hrs at 25ºC ±3ºC
temperature, to prevent battery damage.