DEFIBRILLATOR MONITOR
Mod. 3850B-BIPHASIC - USER MANUAL
Page 45 of 63
WARNING! Verify the AC frequency indicated on the back panel matches the AC frequency which the device is to be connected. If does
not match, contact
feas
ELECTRÓNICA
’s Service Customer. Do not use the device in those conditions. The ECG curve will be
affected by noise and will not be able to use SYNC mode.
WARNING! You must make sure that the AC inlet, to which you will plug the device, has a ground line and is in good condition.
WARNING! The equipment must be connected to an approved electrical installation that includes a correct grounding according to the
local legislation in force. Do not use adapters or replace the device’s original cables. If the plug does not match the installation,
please contact our Customer Service for the provision of a suitable cable.
WARNING! Do not connect this device to an outlet controlled by a switch on/off.
In mobile units, you can power the Monitor Defibrillator from the lighter socket of the unit, through the external 12Vdc power cable.
To connect the device, follow these steps:
1)
Connect the output jack 12Vdc power cable on the back panel.
2)
Connect the plug on the lighter socket.
3)
Verify that the connection is correct and the energy reaches the device, watching green triangular light over battery symbol,
located on the front panel.
4.7. VERIFICATION OF ENERGY DELIVERED
Verification or energy test ensures that the equipment is able to deliver energy on paddles or determine that, due to a fault, does not
allow delivery of energy on paddles. If the test failed, contact the
feas
ELECTRÓNICA
´s Customer Service to arrange repair.
ATTENTION! It is suggested to perform the Test of Energy Delivered according to the domestic policy of the Institution Health or at least
once a week.
ATTENTION! Before performing the Test of Energy Delivered, external paddles must be connected and place on the paddles holder.
Perform the verification with the equipment connected to line power.
WARNING! More than 10 energy discharges should not be performed during Test of Energy Delivered; considering that the minimum
waiting time is 1 minute, between discharges.
For verification of energy delivered perform the following steps:
i.
Turn on the equipment, while pressing the DISARM key, then the equipment will start in energy delivered test mode and you
should see the following message on screen:
ii.
Place the Selector Switch in 200 Joules, as shown in figure above.
iii.
When the Selector Switch is in 200 Joules, and the external paddles are connected to the equipment and placed on the paddles
holder, as indicated in the message of figure above, press the CHARGE key. Then begins a count that is increased up to the value
of about 200 Joules, this number indicates the value of the stored energy, and then the following message is showed in the screen:
iv.
Press the SHOCK button. If the verification result is correct, which means that energy was supplied to the paddles, then you will
see the following message on screen: