DEFIBRILLATOR MONITOR
Mod. 3850B-BIPHASIC - USER MANUAL
Page 39 of 63
Power fault
Indicates a power supply failure in the Oximeter module.
Sensor fault
Indicates that there is a problem with the Oximetry sensor.
Interference by
ambient light
When interference is detected by ambient light, the message "Light Ambient".
3.6.5. CALIBRATION
The functional testers are not calibrated; however an annual contrasting of the oximetry module with a calibrated simulator is suggested.
To do this can send the device to
feas
ELECTRÓNICA
or connect it to a simulator FLUKE-Index2 SpO
2
Simulator. If the set (SpO
2
SpO
2
adapter cable) does not meet the specified tolerances, directly connected to the oximeter, replace the sensor for a new
one. If it still does not meet specifications, send the device to Technical Service.
The following table shows the values of measurements and tolerances.
Oximeter FINAL TEST
ITEM
DESCRIPTION
TO
SIMULATE
TO MEASURE OR
TO CONNECT IN
ADJUST
TARGET
VALUE
MIN MAX
O
XIM
ET
ER
96
Oximetry: Simulate 76% SpO
2
- 40ppm.
CHECK = OK.
-- --
--
--
--
--
97 Register Pulse Rate
--
--
--
40ppm
37
43
98 Register SpO
2
--
--
--
76%
74
78
99
Look at the screen. If shows
plethysmographic curve = OK
-- --
--
--
--
--
100
Oximetry: Simulate 96% SpO
2
- 40ppm.
CHECK = OK.
-- --
--
--
--
--
101 Register Pulse Rate
--
--
--
40ppm
37
43
102 Register SpO
2
--
--
--
96%
94
98
103
Oximetry: Simulate 76% SpO
2
- 240ppm.
CHECK = OK.
-- --
--
--
--
--
104 Register Pulse Rate
--
--
--
240ppm
237
243
105 Register SpO
2
--
--
--
76%
74
78
106 Oximetry: Simulate 96% SpO
2
- 240ppm.
CHECK = OK.
-- --
--
--
--
--
107 Register Pulse Rate
--
--
--
240ppm
237
243
108 Register SpO
2
--
--
--
96%
94
98
109
Simulate "Low Perfusion" (with Cardiosat100,
AMP=4%; with Index2, AMP=0,2%)
(95% SpO2 - 75ppm) CHECK = OK
-- --
--
--
--
--