DEFIBRILLATOR MONITOR
Mod. 3850B-BIPHASIC - USER MANUAL
Page 51 of 63
6. TECHNICAL SPECIFICATIONS
6.1. GENERAL
Device Class I and internally powered.
Degree of protection against electrical shock: Type CF.
Degree of protection against liquids: IPX2.
The equipment is not suitable for use in the presence of explosive atmosphere or flammable mixture.
• Safety Standards: - EN-60601-1 (Eq. IRAM 4220-1 and IEC 601-1).
- EN-60601-1-2. (Electromagnetic Compatibility).
- EN-60601-1-6.
- EN-60601-1-8.
- EN-60601-2-4.
- EN-60601-2-27.
- EN-60601-2-49.
- ISO 80601-2-61.
• Quality Standards: - CE marking, Dir. 93/42/CEE (certificate).
- ISO13485 (certificate).
- A.N.M.A.T. 3266/2013 Disp.: B.P.F. (certificate).
- A.N.M.A.T. 2319/02 Disp. (certificate).
DT: Eng. Jorge F. Feas - MP: 12341991.
The equipment is authorized by A.N.M.A.T. PM-1125-15.
The equipment is sold exclusively to medical professionals and medical institutions.
Use Mode:
Monitoring, Recording and Pacer: continuous use.
Defibrillator: intermittent use, maximum cadence of 200J shock: 3 shock per minute and 1 minute rest.
In both cases, the device will be powered with 90Vac - 240Vac, 50Hz/60Hz.
Electrical Specifications
Power supply: 90Vac - 240Vac (automatic selection) and notch filter 50Hz or 60Hz, factory selectable.
External Power: 12Vdc.
Internal Power Supply
-
Battery.
-
12V-3500mAh Pack.
-
Weight: 750grs.
-
Ni-Mh battery.
Battery Autonomy: Continuous Monitoring 8hs or at least 150 shocks of 200J.
Battery Charging Time from exhausted till at least 90%: 2hrs 15min.
Maximum power consumption during maximum energy charge: 180W.
Power consumption during standby: 7W.
Power supply in Ambulance
- Be medical grade and meet the requirements of standard IEC 60601-1 or come from a vehicle battery (cigarette lighter
connector ambulance).
- Voltage: 12Vdc.
- Maximum Current: 10A.
Line fuse
-
Quantity: 2 (two).
-
Dimensions: 20mm x 5mm.
-
Type: Slow Blow.
-
Voltage: 250V.
-
Current: 2,5A (I
nom
).
-
Speed: 100%
of
I
nom
for 4 hours minimum.
135% of I
nom
for 1 hour maximum.
200% of I
nom
for 120 seconds maximum.
-
Breaking capacity: 100A to 250Vac.
+12V external fuse
-
Dimensions: 20mm x 5mm.
-
Type: Slow Blow.
-
Voltage: 250V.
-
Current: 10A (I
nom
).
-
Speed: 100%
of
I
nom
for 4 hours minimum.
135% of I
nom
for 1 hour maximum.
200% of I
nom
for 120 seconds maximum.
-
Breaking capacity: 100A to 250Vac.
Internal Battery fuse
-
Dimensions: 20mm x 5mm.
-
Type: Slow Blow.
-
Voltage: 250V.
-
Current: 10A (I
nom
).
-
Speed: 100%
of
I
nom
for 4 hours minimum.