DEFIBRILLATOR MONITOR
Mod. 3850B-BIPHASIC - USER MANUAL
Page 22 of 63
•
Use the equipment on a flat and stable surface.
•
Important!! If using a bracket, make sure that the bracket holds at least twice the weight of the equipment. If you have any questions,
please contact
feas
ELECTRÓNICA
.
•
Do not push the keys of the frontal panel with pushing or slicing elements. This will produce permanent damage to the keypad. Only
push the keys of the frontal panel with your fingers. Do not press the buttons with your nails.
Opening the package and verification
Remove the device and its accessories from the packaging carefully and properly keep packaging materials in case of future
transportation or storage.
Check the accessories with the list of accessories purchased. Check if there is any mechanical damage, check the status of all cables
and plug some accessories to review its operation, in case of any problems observed please contact immediately to
feas
ELECTRÓNICA
’s Customer Service.
Installing the device
When installing the device makes sure that the distance from to the wall is 5cm or more, for adequate ventilation.
If the device is installed in a cubicle, you must ensure that the distance between the equipment and any of the walls is 10cm or more.
The device must be supported on a surface capable of supporting twice its weight, with an angle no greater than 10º inclination.
The company guarantees that the device will operate properly and meet all specifications, only if properly installed, as is described in this
User Manual.
2.2. CONNECTIONS AND TURNING ON THE DEVICE
Before connecting any cables, ensure the value of the mains voltage is within the voltage range of the device, also check that the power
frequency indicated on the back panel matches the network’s frequency that will connect the device, if it is different please contact
immediately with
feas
ELECTRÓNICA
’s Customer Service. Do not use the device in these conditions, the ECG curve will be affected
by noise and cannot use the SYNC mode.
Also verify that the output jack of the power cord matches the wall power outlet to which it will connect and possesses electrical energy.
WARINING! The equipment must be connected to an approved electrical installation that includes a correct grounding according to the
local legislation in force. Do not use adapters or replace the device’s original cables. If the plug does not match the installation, please
contact our Customer Service for the provision of a suitable cable.
WARINING! You must make sure that the AC outlet, to which you will plug the device, has a groundling and is in good condition.
WARINING! Do not connect this device to an outlet controlled by a switch on/off.
WARINING! Verify the AC voltage range matches the voltage at which the equipment is to be connected. If does not match, contact
feas
ELECTRÓNICA
’s Service Customer.
WARINING! Verify the AC frequency indicated on the back panel matches the AC frequency which the device is to be connected. If does
not match, contact
feas
ELECTRÓNICA
’s Service Customer. Do not use the device in those conditions. The ECG curve will be
affected by noise and will not be able to use SYNC mode.
WARINING! If you have any doubts about the integrity of the ground, either the cable or the installation of the building, use the device
from the internal battery. In case that the battery is discharge or damage, don't use the equipment.
WARINING! The multi-outlet power extension cord shall not be placed on the floor.
WARINING! The multi-outlet power extension cord should only be used to power equipment that is part of the system.
Once completed these checks, connect the input jack of the power cord to the device, firmly pressing the jack until it stops at the bottom
of the device connector. Then connect the output jack to the wall outlet. Verify that the connection is correct by observing the green
triangular light “Battery Charge”.
Do not attempt to operate this device unless you are thoroughly familiar with these operating
instructions and the function of all controls, indicators, connectors, and accessories.
Turn on the device by turning the selector switch to the MONITOR position (if you only want to use this function) or to the position of
some energy value (if you want to defibrillate). The MONITOR is set initially with ECG input PADDLES (if you do not have the cable
connected to the patient, if you have the cable connected to the patient will start showing the curve of the lead I (DI) on display), OFF
ALARMS (except the lower limit of SpO
2
alarm which defaults to 85%), 25mm/s in sweep speed and 1V/cm in gain.
“Setup error” This message is displayed when, turning on the device, is located that there are differences
between the stored configuration of factory and the current configuration.
The defibrillator has three power modes:
•
90Vac - 240Vac (automatic selection) and notch filter 50Hz or 60Hz, factory selectable.
•
Internal rechargeable battery 12Vdc.
•
External 12Vdc battery.