-47-
After executing defrosting (forced manual defrosting) or
maintenance and checking procedures, it is useful to learn
the next timing by resetting the values.
※
Factory default value
:
0 hour
※
In total operation hours display, [Alarm] lamp
lights up when setting time elapses and indicator
displays [Addt] and measured temperature
alternately, and the unit is under control.
Displaying total operation hours and resetting
1)Press [Set] key for longer than 5 sec. while the unit is
controlled to enter into adjustment mode. Then, the
indicator displays [SnS] and the mode changes to
calibrating displayed temperature mode.
・「
Temperature display lamp
(℃)」
turns off a
light.
2)Press
[▲]
key 6 times to enter into total operation
hours display and reset mode
[
Addt
]
.
3)Press
[
Set
]
key.
Indicator changes to display total operation hour
(blinking).
※
Total operation hours describes the elapsed time
after starting running the unit. Factory default value
is
「
0hr.
」
4)To reset the total operation hour, press
[▲]
key.
Then, the indicator changes to display
「
cLr
」
(
blinking).
[
Total operation hours
]
[▲]↓ ↑[▼]
[
cLr
] :
Reset can be executed.
※
If you do not need to reset, press
[▼]
key to
change the display of indicator to total operation
hours display (blinking) and press [Set] key for
longer than 5 sec. to display measured
temperature. Then, resetting can not be executed.
5) Pressing [Set] key for longer than 5 sec. while
「
cLr
」
is
displayed can display measured temperature on the
indicator and the values can be reset.
※
Values will not be reset until you press
[
Set
]
key for
longer than 5 sec.
※
To change other adjustment mode, press
[▲]
or
[▼]
key to display the mode you' d like to change.
※
If you don't need to change the other adjustment
mode, press [Set] lye for longer than 5 sec. Then,
indicator changes to display [Measured temperature].
※
If no key operation is done for longer than a minute
while setting, the display returns to measured
temperature. In such a case, please reset the setting
1) Press[Set] key for longer than
5 sec. to enter into adjustment
mode.
2)Moves to total
operation hours display and
resets
3)Total operation hours display
4)Resetting total operation hours display
5)Values are reset and measured
temperature is displayed.
7.
How to operate total operation hours display
・
reset
PV
Run
Stop
Mode
Set
▼
▲
Def
h.m
SV
℃
Alarm
Heating
Auto stop
Auto start
1Step
Defrost
24Hour
Pro.
○
○
○
○
○
○
○
○
○
SnS
●
Press 6 times
PV
Run
Stop
Mode
Set
▼
▲
Def
h.m
SV
℃
Alarm
Heating
Auto stop
Auto start
1Step
Defrost
24Hour
Pro.
○
○
○
○
○
○
○
○
○
Addt
●
PV
Run
Stop
Mode
Set
▼
▲
Def
h.m
SV
℃
Alarm
Heating
Auto stop
Auto start
1Step
Defrost
24Hour
Pro.
○
○
○
○
○
○
○
○
1500
●
●
PV
Run
Stop
Mode
Set
▼
▲
Def
h.m
SV
℃
Alarm
Heating
Auto stop
Auto start
1Step
Defrost
24Hour
Pro.
○
○
○
○
○
○
○
○
cLr
●
●
Press 5 times
PV
Run
Stop
Mode
Set
▼
▲
Def
h.m
SV
℃
Alarm
Heating
Auto stop
Auto start
1Step
Defrost
24Hour
Pro.
○
○
○
○
○
○
○
○
○
23.5
●