-17-
Reasons why alarm is activated.
Alarm display and operations
Alarm name
Over load
alarm for
refrigeration
unit
Operating conditions
Refrigeration unit performs over load operation (or
started up).
Releasing alarm
・
Please refer to "Troubleshooting" on page 49.
・
Control will be continued.
・
Buzzer beeps for 15 seconds.
・
Alarm display
Watch dog
・
Resets microcomputer and tries to
restart control. In case that
microcomputer does not get back to
normal operation, control gets
stopped.
・
Alarm display (Not specified).
Operating conditions
Due to noise and etc, temperature controller
is in abnormal condition. Ambient temperature
exceeds 35
℃
.
Releasing alarm
Change the ambience and power source.
Over heat
alarm
・
Stops control of heater.
(
Recovers automatically when
sorting out the trouble).
・
Alarm display
Over Heat
※
Over heat alarm lamp lights up.
Indicator displays measured temperature,
however, other alarm may be activated.
Operating conditions
・
Setting temperature of over heat protector is
low.
・
Temperature controller, SSR or the fan in
the air jacket is impaired, which raises the
setting temperature of over heat protector.
・
Temperature exceeds the settable range of
over heat protector, which made over heat
protecting thermostat work.
(
temperature
reaches around 70
℃
.
)
Over cool
alarm
・
Refrigeration unit will not be controlled.
(
Recovers automatically when sorting
out the trouble).
・
Alarm display
Over Cool
※
Over cool alarm lamp lights up.
Indicator displays measured
temperature,
however, other alarm may be activated.
Operating conditions
・
Setting temperature of over heat protector is
high.
・
Temperature controller, SSR or the fan in
the air jacket is impaired, which lowers the
setting temperature of over cool protector.
Warning function
Reasons why alarm is activated.
Warning display and operations
Warning
Warning for
elapsing total
operation hours
Operating condition
・
Total operation hours exceed the setting time.
Releasing warning
・
Pressing [Set] key can release the warning.
・
Reset total operation hours
「
Addt
」
of
adjustment mode.
※
If you've released the warning by [Set] key
and restart control without resetting total
operation hours, warning will be reactivated
(when turning on and off the unit.)
・
The unit will be controlled.
・