-21-
1.Setting shelf for shaker
(
1
)
Romove shelves and supporters.
(
2
)
Set the supporters from the lowest port of the pole
as appropriate.
(
3
)
Set the shelves board on the supporters and fix
them with screws.
※
Romove supplied ahelf boards and supporters
when using shaker
。
※
Set the shelf board from the lowest step up to the
3rd port from the bottom. Or theunit may
havemalfunction because of the shaking of the
shaker.
※
Set the temperature in the chamber within the
ranger of available ambient temperature of the
product.
Supporter for
shelf (left side)
Shelf
Supporter for
shelf (right
side)
Screw(4)
Modle
Maximum
shakeing speed
MMS-110
MMS-210
MMS-310
MMS-510
150rpm
4-5 Setting optional accessories
Shelf for shaker is used when using shaker,
rotating incubator and low speed magnetic strirrer
in the chamber.
Combination products
(
1
)
Shaker modle MMS-110
・
210
・
310
・
510
※
For details about available shaker, please refer
to the table below.
※
LTI-400E does not accommodate the flask
that has capacity more than 1L.
(
2
)
Rotating incubator Modle MBS-1A
・
1B
※
not available for LTI-400E
(
3
)
Low-speed magnetic striier
model MCS-100
・
101
・
201
※
Shelf boards for LTI-1200E have no port to
insert the power plug of combination products
※
The application of combination products may
be limite depend on the model.
Heat load in the chamber
・
Since the products used in the chamber may
form dew condensation, please prepare water-
proof products.
・
When the heat load of sample or device in
the chamber is too heavy, the temperature may
not reach at the setting value.
Maximum load
0.5LFlaskx4
1LFlaskx2
3LFlaskx1
Weight 5kg
※
When using model LTI-400E
・
700E, outlet(Max.2A)
at the bottom of the chamber is not available while
setting the unit at the bottom of the shelf board.
In such a case, use power plug from the unit installed
from unit set at the cable port and connect the unit to
the power source outside the chamber
Outlet (Max. 1A) on the top of the chamber is used
for fluorescent lamp (optional accessory), however, it
can be connected to other device while the
fluorescent lamp is not being used. In such a case,
sudden surge of electricity may occur when turning the
device on. So check instantaneous largest current as
well as rated current.
※
Capacity of the outlet outside of the chamber of
LTI-400E and the outlet in the chamber of LTI-1200E
is Max. 2A.
※
Since the outlet of LTI has three-terminal, do not
use setting adapter.
Set the shelves for shaker
from the bottom to the 3rd
port from the bottom as
appropriate.
Shelf boards for LTI-
1200E have no port to
insert the power plug of
combination products.