5. Buzzer setting
Buzzer can be beeped when the unit is stopped
automatically in auto stop mode.
※
Selectable items
oFF : NO buzzer
on : Buzzer beeps
※
Factory default
:
oFF
6. Time setting for displaying total operation hours
If you set the time setting for total operation
hours function,
「
Alarm
」
lamp will light up when
the time elapses setting time. Indicator displays
「
Addt
」
and measured temperature alternately
and the unit will be controlled.
※
Setting range of displaying total operation
hours
:
1
〜
9999hrs.
Factory default
:
1500 hrs.
※
Useful to check the timing of executing
defrosting (forced manual defrosting) or
maintenance and checkup.
7. Resetting total operation hours
If you reset the time, you can check the next
timing of defrosting (forced manual defrosting),
maintenance and checkup after resetting the time.
※
Factory default
:
0hr.
The setting of adjustment mode can be changed even
when the unit is controlled or not controlled.
1) Press
[
Set
]
key for longer than 5 seconds.
Indicator displays
「
SnS
」
and the mode enters
into adjustment mode.
2) Each time you press
[▲]
or
[▼]
key, setting
mode
changes. Choose your desired mode.
※
Adjustment mode changes in order as shown in
right picture.
For more details about operation, please refer to "How
to operate adjustment mode" on page 41.
[▼]
[▲]「
SnS
」 :
Calibrating displayed
temperature
↓ ↑ [▼]
[▲]「
AL-H
」 :
Upper temperature limit
alarm
↓ ↑ [▼]
[▲] 「
AL-L
」 :
Lower temperature limit
alarm
↓ ↑ [▼]
[▲] 「
PoFF
」 :
Power recovery
setting
↓ ↑ [▼]
[▲] 「
buSE
」 :
Buzzer setting
↓ ↑ [▼]
[▲] 「
Adtt
」 :
Displaying total operation
hour setting
↓ ↑ [▼]
[▲]
「
Addt
」 :
Displaying total
operation hours
・
reset
-14-
※
If you set the buzzer setting, buzzer will keep on
beeping when auto stop control will be
completed (when stopping control).
Pressing [Set] key can stop the buzzer.