- 3 -
Model
Overall dimensions *3
Power input
Rated power source
Net weight *4
5
~
35
℃
15A 1.5kVA
AC100V 50/60Hz
Range of available temperature
635W
×
755D
×
1695H mm
153
㎏
KCL-2000W
Recorder output
Measured temperature : DC1mV/
℃
Measured humidity : DC1mV1
%
RH
Alarm output
Contact capacity AC250V 5A Normal : OFF Abnormal : ON
Drain port
Nozzle OD 11
×
ID7 mm
*1
*2
*3
*4
*
The above performance has been obtained under conditions of 20
℃
of room temperature, rated power
source voltage, and without load.
The performance is indicated
in conformity with JTM KO1-1998 (Testing method and indication form
for constant temperature and humidity bath by Japan Testing Machinery Association).
If the unit is operated at 30
~
40
℃
or lower temperature for a long time (one day
~
one week or longer),
please set “interval defrosting operation” for the operation.
The chamber temperature will go up by about 3
℃
, and the chamber humidity increase by about
20%RH, when the defrosting mode works under room temperature 20
℃
.
Actual changes of such temperature and humidity differ depend on actual operating conditions.
The chamber dimensions and overall dimensions without external tank do not include any projections.
The net weight includes external tank part and rack frame.
If the rack is not used, the net weight is about 127kg.
The required time to heat and cool, and to humidify, temperature and humidity control accuracy, and
temperature and humidity uniformity doffer depend on the environmental temperature, power source
voltage, with load or without load.