- 24 -
2. Setting shelves
Check which bracket is for right and for left.
The bracket must be attached with setting
the down-turned hook into the front side
bracket holder and the sideways hook into
the back side bracket holder.
First inset the sideways hook into a hole of
the bracket holder, and push it backward
surely.
Next inset the down-turned hook into a hole
of the front side of bracket holder, and push
it downward surely.
Make sure that each hole level is even
between the front and the back.
After setting all shelves, check the level of
each face-to-face bracket is even to set
shelves horizontally.
Put shelves on each bracket.
(1)
(2)
(3)
(4)
(5)
(6)
3. Preparation of water-supply
Bracket holder
Sideways hook
Bracket (Left)
Down-turned hook
2
1
Socket
Nut for socket
Hose (supplied)
Replacement of hose of water supplying hose set
Put the hose into the nut for socket,
connect it to the socket, and tighten the nut.
Connect the L-shaped socket to the
another end of hose, and fix it with a hose
band.
* When attaching the tank to the left side of unit,
replace the hose of water-supply hose set with the
supplied another hose.
Make sure that no leakage is found from connecting
parts after replacement.
Caution
Use purified water such as ion-
exchange water or distilled
water for humidifying.
Non-purified water pollutes humidification
heater, wet-bulb wick, and tubings. It obstructs
the normal humidification control. And it may
occur water leakage or other troubles.
Rack for
external
water
supply tank
External water supply tank
Supply purified water into the supplied
external tank which electric conductivity is
10
µ
s/cm or less. The tank has 20L capacity.
Mount the tank on the rack.
(1)
(2)