- 20 -
Removing rack and installation without rack
(1)
(2)
The unit is fixed to the rack with
hexagon
head bolts
and nuts (4 parts).
Loosen hexagon head bolts and nuts with
using two spanners (
opposite side
distance : 13).
Move the unit gently to the installation site.
Adjust four level adjusters to make the unit
horizontal with using a spanner
(
opposite
side distance : 17).
* This adjustment is temporary.
4-4 Installation --- Adjustment of level ---
Make sure the position where
external water-supply tank is set up
before installing the unit.
When you lift or move the unit, do
not hold the ventilation grill part to
prevent falling down the unit.
Rack
Level adjuster (4 pcs.)
Ventilation cover
Rotation part
Lower
Rise
Level adjuster
Hexagon
head bolt
Hexagon
head nut
Spring
washer
Flat washer
Flat washer
Make sure that the unit is set up in a
horizontal position.
Caution
Put spirit level on the top of the unit to
make sure that the it is in a horizontal
position. When the unit is not horizontal,
adjust four adjusters either of unit or rack
with using a
spanner
(
opposite side
distance : 17) .
If the unit is not horizontal, the wick holder
may not be supplied the water during operation,
or supplied water may over flow from the wick
holder or from the humidification tray (top/bottom),
which may increase the purified water consumption.
If you cannot prepare a spirit level, you
can adjust four adjusters
temporarily
to make the filled water surface of the
bottom humidification tray horizontal.
Refer to [Humidification procedure] on
P.33 to know more details.