7. Using cable port (Connection of monitoring devices)
The measured temperature is output DC1mV per 1
℃
.
Output accuracy is
±
2.6
℃
of the measured temperature.
<<A case of output>>
Measured temperature : -15.0
℃
→
Recorder output : -15.0
±
2.6 mV
Measured temperature : 85.0
℃
→
Recorder output : 85.0
±
2.6 mV
- Recorder (Analog) output terminal for measured temperature
Temperature
Humidity
+
-
G
(Ground)
+
-
G
(Ground)
Inset
From a recorder
Banana-shape jack
To a recorder
Lead wire
Temperature
Humidity Alm.Out
Com
Alarm output terminal
RS-232C interface
Left view of unit
Cable port
(one port each for both right and left side of unit)
Use the port depending on your purpose.
The cable port (one port each for both right
and left side of unit) can be used to connect
some monitoring tools or to connect cables
of inner devices to outer electric power
supply port.
Embed a supplied silicone stopper into the
cable port which is made a lengthwise cut
in. It prevent entering moisture into the
chamber. The silicone stopper is supplied
one piece as standard.
8. Connection of recorder (Analog) output terminal
Use the terminal depending on your purpose.
Recorder output
terminal for
temperature
Recorder output
terminal for
humidity
Inset
The measured humidity is output DC1mV per 1%RH.
Output accuracy is
±
2.0%RH of the measured humidity.
<<A case of output>>
Measured humidity : 25.0%RH
→
Recorder output : 25.0
±
2.0 mV
Measured humidity : 85.0%RH
→
Recorder output : 85.0
±
2.0 mV
- Recorder (Analog) output terminal for measured humidity
9. Connection to RS-232C interface
Use the interface depending on your purpose.
The unit can communicate with a host
computer through the RS-232C interface.
The communication speed is 9600BPS as
fixed. Refer to the page 49 and 50 of the
key operation manual.