- 44 -
Description
Wet-bulb wick
Quantity
1 dozen
Cat. No.
205190
Take out the filter from the ventilation
cover.
Tap the filter to remove dust, and
wash it with water.
(Use some neutral detergent, if
needed.)
Dry the filter enough.
(Since the filter is made of heat-
sensitive material, do not use dryer
and etc .)
Attach the filter to the ventilation
cover, and close it.
(2)
(3)
(4)
(5)
Condenser (Cooling fin)
Filter
Observation window
for humidification tray
(top)
Wick
holder
Observation window
for attached frost
Suction port cover
Heater for
humidification
humidification tray
(top)
Overflow port
Overflow port
Heater for
humidification
humidification tray
(bottom)
4. Draining from wick holder and humidification vat (top/bottom)
If you use the unit for a prolonged
period, water stain and scale or some
algae attach. Press [Manual Drain]
key to run the forced draining
operation, and change the water
during periodic intervals.
The water is drained completely for
about 5 minutes later.
5. Cleaning of humidification trays (top/bottom)
You can find the top humidification tray
through the observation window.
And the bottom humidification tray is put
in the back of the suction port cover.
Clean both humidification trays regularly.
Turn on the [Manual Drain] key to
drain the water of wick holder and
both humidification trays.
Turn off the power switch and the
mains switch.
Loosen four screws of the second step
plate from the top of rear side, which
has the observation window.
Hold the bottom edge of suction port
cover, and pull it to detach.
Clean both top and bottom
humidification vat. As each tray has a
strainer, remove it to clean.
You can ask the maintenance of unit
to your service agent at regular
intervals.
(1)
(2)
(3)
(4)
(5)
*
6. Check-up of wet-bulb wick
The wet-bulb wick is a consumable
part accessory.
Check it as needed depending on the
operating condition. If it is discolored
or deteriorated, replace it with new
one.
Change the wick every month .