- 30 -
10. Connection of alarm output terminal
Drain Tank Alm.In
Use the terminal depending on your purpose.
11. Connection of an optional drain tank
Use the drain tank depending on your purpose.
12. Connection of mains plug
Alarm output connector
Terminal
Lead wire
Thermal contraction tube
Soft-soldering
* Soft-solder lead wires, and contract thermal
contraction tubes with a hair dryer.
Water supply plug
Drain nozzle
Alarm input terminal
for drain tank
To alarm input terminal
for drain tank
Connector
Hose
To drain nozzle
Socket
Cap
Socket
Push the socket in the direction
of an arrow to disconnect.
When a safety function is active, the terminal
outputs and indicates the trouble of the unit at
contact point. The contact point capacity is
AC250V, 5A with resistant load. When the
unit is in abnormal status, it is turned ON (It
is turned OFF in normal status).
Disassemble the supplied alarm output
connector, and soft-solder the lead wires of
external device to each terminal as shown
on the right figure.
To avoid short circuit between both
terminals, insulate the wire with each
thermal contraction tube.
Do not use without soft-soldering.
Assemble the connector again.
Connect the alarm output plug to the port
[Alm Out].
(1)
*
*
(2)
(3)
The capacity of drain tank is 10L, and it
collects the drained water from the chamber.
When the tank is full, the alarm [DTK] is
indicated on the LCD panel of unit, and the
humidification is stopped.
Drain the water from the tank before the water
level exceeds 10L on the gauge.
Hold the tank with the rack, connect the
connector to the alarm input terminal, and
connect the hose to the drain nozzle.
Do not twist the hose or do not hang lower
than the inlet of drain tank. The water may
not be drained to the tank because of piping
resistance, or overflow from the chamber.
Cut the hose proper length in accordance
with the place of the tank.
*
When draining water from the tank, remove the
socket of drain tank hose. If you would like to drain
the water from the unit, drain it out into a tray.
Disconnect the input terminal connector, remove the
cap and drain the water.
Push the black part of socket in the
direction of an arrow to disconnect it. To
connect again, inset the socket part until
you hear a click.
*
Make sure that the power switch and the mains switch are turned off, and disconnect the mains
connector from AC outlet port.