Suspend and Resume
394
FTB-8100 Series Transport Blazer
Resume Operation
Resume Operation
To enter the resume operation, turn on the unit by pressing the On/Off
button. During the resume operation, the module is re-initialized and once
its booting cycle is completed the test is created with the configuration
maintained in static RAM.
In the event a power failure or manual shutdown of the FTB-200 occurs,
while the resume operation is in-progress, the resume operation is
terminated and all the test configuration information is lost.
The following conditions prevent the resume operation:
³
The module status has changed (mismatch in serial number or
module is missing).
³
A module failure is detected.