SONET Tabs
SONET/SDH Application
177
Line TX (SONET)
For
Manual
method:
³
Amount
:
Select the amount of manual error to be generated. Choices
are
1
through
50
. The default setting is
1
.
³
Send
button:
Press
Send
to manually generate error(s) according to
the
Error Type
and the
Amount of Errors
selected.
For
Rate
method:
³
Rate
:
Select the injection rate for the selected error. The rate must be
within the minimum and maximum values specified.
³
Continuous
:
Generates the selected error to its theoretical maximum
rate when the
Continuous
check box is selected. The
Continuous
check box is cleared by default.
³
On/Off
button: The On/Off button is used to activate/deactivate the
selected error at the rate specified or at its theoretical maximum rate
when the
Continuous
check box is selected. This setting is disabled
(Off) by default.
Alarm Generation
Allows
Continuous
alarm generation method.
³
Type
AIS-L
(Alarm Indication Signal - Line): Generates a SONET signal that
contains a valid Section Overthead (SOH) and an all-ones pattern on
the SPE.
RDI-L
(Remote Defect Indication - Line): Generates a “110” pattern for
the bits 6, 7 and 8 of the K2 byte.
The default setting is
AIS-L
³
On/Off
button: The On/Off button is used to activate/deactivate the
selected alarm. This setting is disabled (Off) by default.
Содержание FTB-8100 Series
Страница 1: ...User Guide FTB 8100 Series for FTB 200 SONET SDH Application...
Страница 18: ......
Страница 22: ......
Страница 24: ......
Страница 94: ......
Страница 108: ......
Страница 178: ......
Страница 224: ......
Страница 260: ......
Страница 312: ......
Страница 352: ......
Страница 382: ......
Страница 412: ......
Страница 490: ......
Страница 500: ......