S10k
Description
3.
13
3. Description
3.1
Design
The system consists of the following main components:
•
The separately mounted injector (6) for the generation of the
operating vacuum and for the mixing of the gas with the water.
•
The vacuum demand valve (1) is connected directly to the Cl
2
or SO
2
gas cylinder or to the gas manifold. It only opens when
the vacuum created by the injector is sufficient.
For the use with carbon dioxide a pressure reducing valve to
13.5 bar (g) is necessary.
•
The gas dosing unit (2) connected to the vacuum demand
valve via the vacuum line.
It includes the flowmeter to display the gas flow and the V-
notch plug to adjust the gas flow.
Dosage can be switched on and off by selenoid valves in the water
supply line or by booster pump.
Содержание WALLACE & TIERNAN S10k
Страница 1: ...WALLACE TIERNAN GAS FEEDER S10k INSTRUCTION MANUAL ...
Страница 2: ...2 WT 025 200 000 DE IM 0517 S10k Please note Original manual ...
Страница 6: ...6 WT 025 200 000 DE IM 0517 Contents S10k ...
Страница 22: ...22 WT 025 200 000 DE IM 0517 Description S10k 3 ...
Страница 88: ...88 WT 025 200 000 DE IM 0517 Drawings S10k 7 7 4 Spare parts for the injectors 7 4 1 Injector W3T171367 1 ...
Страница 90: ...90 WT 025 200 000 DE IM 0517 Drawings S10k 7 7 4 2 Anti syphon injector W3T171368 1 ...
Страница 93: ...S10k Drawings 7 93 ...
Страница 94: ...94 WT 025 200 000 DE IM 0517 Drawings S10k 7 7 4 3 Injector W3T171369 3 4 ...
Страница 96: ...96 WT 025 200 000 DE IM 0517 Drawings S10k 7 7 4 4 Antisyphon Injektor W3T171370 3 4 ...
Страница 99: ...S10k Manufacturer declaration 8 99 8 Manufacturer declaration ...
Страница 100: ...100 WT 025 200 000 DE IM 0517 Manufacturer declaration S10k 8 ...