34
WT.050.810.040.DE.IM.0519
Installation
DEPOLOX
®
5 E flow cell
4.
4.4.5
Removing or fitting the housing cover
1
Remove the housing cover of the DEPOLOX
®
5 E flow cell. To do
this, press both unlocking buttons on the top of the housing and care
-
fully remove the housing cover forwards.
2
Refit and engage the housing cover of the DEPOLOX
®
5 E flow cell.
To do this, place the housing cover against at the bottom of the hous
-
ing and carefully push it upwards until the housing cover engages
with the unlock buttons.
Image 1 Top view of housing cover
A Unlocking buttons
4.4.6
Connecting the sample water inlet
Please note
Do not install water carrying lines made of copper. These would distort
the measurements.
When connecting the sample water inlet, observe the following:
• Sample water inlets are classified according to whether they have a
hose connection or are fitted with rigid piping.
• The pressure in the sample water inlet must always be within a range
of 0.25 to 3.0 bar. The pressure in the sample water inlet must gener
-
ally be 0.25 bar higher than in the sample water outlet.
• At an admission pressure below 0.25 bar, an optional pressure
booster pump must be used. See chapter 4.4.7 – “Example for sam
ple water take-off with a booster pump”.
• If the admission pressure exceeds 3.0 bar, an optional pressure
reducing valve must be used. See chapter 4.4.7 – “Example for sam
• To prevent long loop lag times, ensure that the lines in the sample
water inlet are as short as possible.
• An external strainer with a mesh width of 0.5 mm is provided for the
sample water inlet.
Image 2 Cross-section of DEPOLOX
®
5 E flow cell, non-pressurized
A Sample water inlet with shut-off ball valve
B Sample taking unit (drain on the drain screw)
C Sample water outlet
A
A
A B C
Содержание WALLACE & TIERNAN DEPOLOX 5 E 700 P Series
Страница 1: ...WALLACE TIERNAN DEPOLOX 5 E FLOW CELL INSTRUCTION MANUAL...
Страница 2: ...2 WT 050 810 040 DE IM 0519 DEPOLOX 5 E flow cell Please note Original instruction manual Simone...
Страница 56: ...56 WT 050 810 040 DE IM 0519 Maintenance DEPOLOX 5 E flow cell 5...
Страница 72: ...72 WT 050 810 040 DE IM 0519 Declarations and certificates DEPOLOX 5 E flow cell 7...
Страница 76: ...76 WT 050 810 040 DE IM 0519 Declarations and certificates DEPOLOX 5 E flow cell 7...
Страница 80: ...80 WT 050 810 040 DE IM 0519 Index DEPOLOX 5 E flow cell 8...